Roland L, Drillich M, Fidlschuster B, Schwendenwein I, Iwersen M
Clinical Unit for Herd Health Management for Ruminants, University Clinic for Ruminants, Department for Farm Animals and Veterinary Public Health, University of Veterinary Medicine Vienna, 1210 Vienna, Austria.
Central Diagnostic Unit, Department for Pathobiology, University of Veterinary Medicine Vienna, 1210 Vienna, Austria.
J Dairy Sci. 2014 Sep;97(9):5580-6. doi: 10.3168/jds.2013-7736. Epub 2014 Jul 3.
The objective of this study was to evaluate the suitability of the V-Sight hematology analyzer (A. Menarini Pharma GmbH, Vienna, Austria) for bovine blood by a comparison with a reference device (Advia 2120i, Siemens AG, Erlangen, Germany). In total, 97 blood samples were obtained from 75 dairy cows. Analyzed parameters included counts of white blood cells (WBC), lymphocytes, monocytes, granulocytes, red blood cells (RBC), and platelets (PLT), as well as hemoglobin concentration (HGB), hematocrit (HCT), mean corpuscular volume, mean corpuscular hemoglobin (MCH), MCH concentration (MCHC), mean platelet volume (MPV), and plateletcrit (PCT). Based on Passing-Bablok regression, the V-Sight provided accurate and precise results for MCH and MCHC only. The PCT results were comparable to the reference method, but precision was inconclusive. Significant proportional differences were detected for monocytes, granulocytes, HCT, and PLT. For all other analytes, significant proportional and systemic differences were observed. The WBC and lymphocyte results from the V-Sight were characterized by poor accuracy, poor precision, and a high number of false positive outliers. Bland-Altman analysis indicated negative biases for all WBC parameters, the erythrocyte indices, and PLT. Positive biases were observed for RBC, HGB, HCT, MPV, and PCT. Correlation coefficients of >0.9 between the V-Sight and the reference method were found only for RBC, HGB, HCT, and MPV. Intraassay precision of the V-Sight analyzer was acceptable (coefficient of variation <5%) for granulocytes, the erythrocyte indices, and MPV. It was unacceptable (coefficient of variation ≥5%) for WBC, lymphocytes, monocytes, as well as RBC, and inconclusive for HGB, HCT, PLT, and PCT. Sensitivity was high for all RBC counts and indices as well as PLT, but low for monocytes, granulocytes, and MPV. Specificity was high for monocytes and granulocytes, but low for RBC, HCT, MCH, and MCHC. With accurate and precise results for only 2 out of 13 parameters, the V-Sight cannot be recommended for analysis of bovine blood.
本研究的目的是通过与参考设备(德国埃尔朗根西门子公司的Advia 2120i)比较,评估V-Sight血液分析仪(奥地利维也纳A. Menarini制药有限公司)对牛血液的适用性。总共从75头奶牛身上采集了97份血液样本。分析的参数包括白细胞(WBC)、淋巴细胞、单核细胞、粒细胞、红细胞(RBC)和血小板(PLT)计数,以及血红蛋白浓度(HGB)、血细胞比容(HCT)、平均红细胞体积、平均红细胞血红蛋白含量(MCH)、MCH浓度(MCHC)、平均血小板体积(MPV)和血小板压积(PCT)。基于Passing-Bablok回归分析,V-Sight仅在MCH和MCHC方面提供了准确且精确的结果。PCT结果与参考方法具有可比性,但精密度尚无定论。在单核细胞、粒细胞、HCT和PLT方面检测到显著的比例差异。对于所有其他分析物,观察到显著的比例差异和系统差异。V-Sight的WBC和淋巴细胞结果准确性差、精密度差且有大量假阳性异常值。Bland-Altman分析表明,所有WBC参数、红细胞指数和PLT均存在负偏差。RBC、HGB、HCT、MPV和PCT存在正偏差。V-Sight与参考方法之间仅在RBC、HGB、HCT和MPV方面发现相关系数>0.9。V-Sight分析仪对粒细胞、红细胞指数和MPV的批内精密度可接受(变异系数<5%)。对WBC、淋巴细胞、单核细胞以及RBC不可接受(变异系数≥5%),对HGB、HCT、PLT和PCT尚无定论。所有RBC计数和指数以及PLT的灵敏度较高,但单核细胞、粒细胞和MPV的灵敏度较低。单核细胞和粒细胞的特异性较高,但RBC、HCT、MCH和MCHC的特异性较低。由于13个参数中只有2个参数的结果准确且精确,因此不推荐使用V-Sight分析牛血液。