To compare the equivalent optic nerve head (OHN) parameters obtained with confocal scanning laser ophthalmoscopy (HRT3) and spectral-domain optical coherence tomography (OCT) in healthy and glaucoma patients.
One hundred and eighty-two consecutive healthy subjects and 156 patients with open-angle glaucoma were divided into 2 groups according to intraocular pressure and visual field outcomes. All participants underwent imaging of the ONH with the HRT3 and the Cirrus OCT. The ONH parameters and the receiver operating characteristic (ROC) curves were compared between both groups.
Mean age did not differ between the normal and glaucoma groups (59.55 ± 9.7 years and 61.05 ± 9.4 years, resp.; P = 0.15). Rim area, average cup-to-disc (C/D) ratio, vertical C/D ratio, and cup volume were different between both instruments (P < 0.001). All equivalent ONH parameters, except disc area, were different between both groups (P < 0.001). The best areas under the ROC curve were observed for vertical C/D ratio (0.980 for OCT and 0.942 for HRT3; P = 0.11). Sensitivities at 95% fixed-specificities of OCT parameters were higher than those of HRT3.
Equivalent ONH parameters of Cirrus OCT and HRT3 are different and cannot be used interchangeably. ONH parameters measured with OCT yielded a slightly better diagnostic performance.
比较共焦激光扫描检眼镜(HRT3)和频域光相干断层扫描(OCT)在正常人和青光眼患者中获得的等效视盘参数。
根据眼压和视野结果,将 182 例连续的健康受试者和 156 例开角型青光眼患者分为 2 组。所有参与者均接受 HRT3 和 Cirrus OCT 的视神经头成像。比较两组之间的视神经头参数和受试者工作特征(ROC)曲线。
正常组和青光眼组的平均年龄无差异(59.55 ± 9.7 岁和 61.05 ± 9.4 岁,P = 0.15)。边缘区域、平均杯盘比、垂直杯盘比和杯容积在两种仪器之间存在差异(P < 0.001)。除视盘面积外,两种仪器的所有等效视盘参数均存在差异(P < 0.001)。ROC 曲线下的最佳面积观察到垂直杯盘比(OCT 为 0.980,HRT3 为 0.942;P = 0.11)。OCT 参数的 95%固定特异性的灵敏度高于 HRT3。
Cirrus OCT 和 HRT3 的等效视盘参数不同,不能互换使用。OCT 测量的视盘参数具有稍好的诊断性能。