Suppr超能文献

使用肌电图/生物阻抗联合测量系统对吞咽进行自动检测和评估。

Automated detection and evaluation of swallowing using a combined EMG/bioimpedance measurement system.

作者信息

Schultheiss Corinna, Schauer Thomas, Nahrstaedt Holger, Seidl Rainer O

机构信息

Ear, Nose and Throat Unit, Unfallkrankenhaus Berlin, Warener Straße 7, 12683 Berlin, Germany.

Control Systems Group, Technische Universität Berlin, Germany.

出版信息

ScientificWorldJournal. 2014;2014:405471. doi: 10.1155/2014/405471. Epub 2014 Jul 10.

Abstract

INTRODUCTION

Developing an automated diagnostic and therapeutic instrument for treating swallowing disorders requires procedures able to reliably detect and evaluate a swallow. We tested a two-stage detection procedure based on a combined electromyography/bioimpedance (EMBI) measurement system. EMBI is able to detect swallows and distinguish them from similar movements in healthy test subjects.

STUDY DESIGN

The study was planned and conducted as a case-control study (EA 1/019/10, and EA1/160/09, EA1/161/09).

METHOD

The study looked at differences in swallowing parameters in general and in the event of penetration during swallows in healthy subjects and in patients with an oropharyngeal swallowing disorder. A two-stage automated swallow detection procedure which used electromyography (EMG) and bioimpedance (BI) to reliably detect swallows was developed.

RESULTS

Statistically significant differences between healthy subjects and patients with a swallowing disorder were found in swallowing parameters previously used to distinguish between swallowing and head movements. Our two-stage algorithm was able to reliably detect swallows (sensitivity = 96.1%, specificity = 97.1%) on the basis of these differences.

DISCUSSION

Using a two-stage detection procedure, the EMBI measurement procedure is able to detect and evaluate swallows automatically and reliably. The two procedures (EMBI + swallow detection) could in future form the basis for automated diagnosis and treatment (stimulation) of swallowing disorders.

摘要

引言

开发一种用于治疗吞咽障碍的自动化诊断和治疗仪器需要能够可靠地检测和评估吞咽的程序。我们测试了一种基于肌电图/生物阻抗(EMBI)组合测量系统的两阶段检测程序。EMBI能够检测吞咽并将其与健康测试对象中的类似运动区分开来。

研究设计

该研究作为病例对照研究进行规划和实施(EA 1/019/10、EA1/160/09、EA1/161/09)。

方法

该研究观察了健康受试者和口咽吞咽障碍患者在一般吞咽参数以及吞咽时发生误吸情况下的差异。开发了一种使用肌电图(EMG)和生物阻抗(BI)来可靠检测吞咽的两阶段自动吞咽检测程序。

结果

在先前用于区分吞咽和头部运动的吞咽参数方面,发现健康受试者与吞咽障碍患者之间存在统计学上的显著差异。基于这些差异,我们的两阶段算法能够可靠地检测吞咽(敏感性 = 96.1%,特异性 = 97.1%)。

讨论

使用两阶段检测程序,EMBI测量程序能够自动且可靠地检测和评估吞咽。这两个程序(EMBI + 吞咽检测)未来可能构成吞咽障碍自动化诊断和治疗(刺激)的基础。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/1622/4121244/ee278086657e/TSWJ2014-405471.001.jpg