使用具有不同翻转角演变的应用优化对比的三维采样完美技术评估垂体微病变。

Assessment of pituitary micro-lesions using 3D sampling perfection with application-optimized contrasts using different flip-angle evolutions.

作者信息

Wang Jing, Wu Yue, Yao Zhenwei, Yang Zhong

机构信息

Department of Radiology, Huashan Hospital, Fudan University, 12 Wulumuqi Road, Shanghai, 200040, People's Republic of China.

出版信息

Neuroradiology. 2014 Dec;56(12):1047-53. doi: 10.1007/s00234-014-1432-1. Epub 2014 Sep 24.

Abstract

INTRODUCTION

The aim of this study was to explore the value of three-dimensional sampling perfection with application-optimized contrasts using different flip-angle evolutions (3D-SPACE) sequence in assessment of pituitary micro-lesions.

METHODS

Coronal 3D-SPACE as well as routine T1- and dynamic contrast-enhanced (DCE) T1-weighted images of the pituitary gland were acquired in 52 patients (48 women and four men; mean age, 32 years; age range, 17-50 years) with clinically suspected pituitary abnormality at 3.0 T, retrospectively. The interobserver agreement of assessment results was analyzed with K-statistics. Qualitative analyses were compared using Wilcoxon signed-rank test.

RESULTS

There was good interobserver agreement of the independent evaluations for 3D-SPACE images (k = 0.892), fair for routine MR images (k = 0.649). At 3.0 T, 3D-SPACE provided significantly better images than routine MR images in terms of the boundary of pituitary gland, definition of pituitary lesions, and overall image quality. The evaluation of pituitary micro-lesions using combined routine and 3D-SPACE MR imaging was superior to that using only routine or 3D-SPACE imaging.

CONCLUSION

The 3D-SPACE sequence can be used for appropriate and successful evaluation of the pituitary gland. We suggest 3D-SPACE sequence to be a powerful supplemental sequence in MR examinations with suspected pituitary micro-lesions.

摘要

引言

本研究旨在探讨采用不同翻转角演变的三维采样完美与应用优化对比(3D-SPACE)序列在评估垂体微病变中的价值。

方法

回顾性地对52例临床怀疑垂体异常的患者(48例女性和4例男性;平均年龄32岁;年龄范围17 - 50岁)在3.0 T条件下采集垂体的冠状位3D-SPACE以及常规T1加权和动态对比增强(DCE)T1加权图像。采用K统计分析评估结果的观察者间一致性。使用Wilcoxon符号秩检验比较定性分析结果。

结果

3D-SPACE图像独立评估的观察者间一致性良好(k = 0.892),常规MR图像的观察者间一致性一般(k = 0.649)。在3.0 T时,3D-SPACE在垂体边界、垂体病变清晰度和整体图像质量方面提供的图像明显优于常规MR图像。联合使用常规和3D-SPACE MR成像对垂体微病变的评估优于仅使用常规或3D-SPACE成像。

结论

3D-SPACE序列可用于对垂体进行恰当且成功的评估。我们建议3D-SPACE序列在怀疑有垂体微病变的MR检查中是一种强大的补充序列。