99锝-巯基乙酰三甘氨酸与131碘-邻碘马尿酸盐在测定有效肾血浆流量(ERPF)中的比较。

Comparison of 99Tcm-mercaptoacetyltriglycine and 131I-orthoiodohippurate in determination of effective renal plasma flow (ERPF).

作者信息

Abdel-Dayem H M, Sadek S, al-Bahar R, Sabha M, el-Sayed M

机构信息

Department of Nuclear Medicine, Faculty of Medicine, Kuwait University, Safat.

出版信息

Nucl Med Commun. 1989 Feb;10(2):99-107. doi: 10.1097/00006231-198902000-00004.

Abstract

Recent studies have suggested that 99Tcm-labelled mercaptoacetyltriglycine (99Tcm-MAG3) is a potential replacement for 131I-orthoiodohippurate (131I-OIH). The two radiopharmaceuticals were studied in 19 patients with varying degrees of renal impairment, for comparison of effective renal plasma flow (ERPF). All patients were injected i.v. with 2-3 mCi 99Tcm-MAG3 and 250-300 microCi of 131I-OIH simultaneously in separate arms to avoid any chemical incompatibility. Serial digital images were acquired simultaneously for 99Tcm-MAG3 and 131I-OIH at 15 s per frame for 30 min. The renogram curve was generated and ERPF value was calculated for each kidney using both radiopharmaceuticals and applying the same computer program (Schlegel's program). In all cases, the 99Tcm-MAG3 and 131I-OIH renogram curves were quite similar. The values of ERPF of 99Tcm-MAG3 were not significantly different from those of 131I-OIH studies (paired t test, p less than 0.01). The correlation coefficient between the two methods was 0.95 (p less than 0.001) with a linear regression equation Y = -12.96 + 0.995X, where Y = ERPF determined by 131I-OIH and X = ERPF determined by 99Tcm-MAG3. This study indicates that 99Tcm-MAG3 is a useful 99Tcm replacement for 131I-OIH in the determination of ERPF and generation of renogram curves.

摘要

近期研究表明,99锝标记的巯基乙酰三甘氨酸(99Tcm-MAG3)可能替代131碘邻碘马尿酸(131I-OIH)。对19例不同程度肾功能损害的患者进行了这两种放射性药物的研究,以比较有效肾血浆流量(ERPF)。所有患者均通过静脉注射,在不同手臂同时分别给予2 - 3毫居里的99Tcm-MAG3和250 - 300微居里的131I-OIH,以避免任何化学不相容性。以每秒15帧的速度同时采集99Tcm-MAG3和131I-OIH的系列数字图像,持续30分钟。生成肾图曲线,并使用这两种放射性药物并应用相同的计算机程序(施莱格尔程序)计算每个肾脏的ERPF值。在所有病例中,99Tcm-MAG3和131I-OIH的肾图曲线非常相似。99Tcm-MAG3的ERPF值与131I-OIH研究的结果无显著差异(配对t检验,p小于0.01)。两种方法之间的相关系数为0.95(p小于0.001),线性回归方程为Y = -12.96 + 0.995X,其中Y = 由131I-OIH测定的ERPF,X = 由99Tcm-MAG3测定的ERPF。本研究表明,在测定ERPF和生成肾图曲线方面,99Tcm-MAG3是一种有用的可替代131I-OIH的99锝标记药物。

相似文献

4
Comparative study of renal scintigraphy with 99Tcm-mercaptoacetyltriglycine and 123I-orthoiodohippurate.
Nucl Med Commun. 1993 Aug;14(8):644-52. doi: 10.1097/00006231-199308000-00004.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验