Michel Adrien, Bosc Romain, Mathieu Vincent, Hernigou Philippe, Haiat Guillaume
CNRS, Laboratoire de Modélisation et de Simulation Multi-Echelle - UMR CNR 8208, Créteil, France.
Service de Chirurgie Orthopédique et Traumatologique, Hôpital Henri Mondor AP-HP, CHU Paris 12, Université Paris-Est, France.
Proc Inst Mech Eng H. 2014 Oct;228(10):1027-34. doi: 10.1177/0954411914552433. Epub 2014 Sep 25.
Press-fit procedures used for the insertion of cementless hip prostheses aim at obtaining optimal implant primary stability. We have previously used the measurement of impact duration to follow the insertion of the acetabular cup implant within bone tissue. The aim of this study was to investigate the variation of the value of the impact momentum due to successive insertions of the acetabular cup into bone tissue. The results obtained with impact momentum and contact duration measurements were compared. A total of 10 bovine bone samples were subjected to three successive procedures consisting of 10 reproducible impacts (3.5 kg falling 40 mm). Each procedure aimed at inserting the acetabular cup implant into the same bone cavity. The time variation of force during each impact was recorded by a force sensor, allowing the measurement of the impact duration (I 1) and momentum (I 2). The value of I 2 increased as a function of the impact number and reached a constant value after N 2 = 5.07 ± 1.31 impacts. Moreover, statistical analyses show that N 2 decreased significantly as a function of the number of experiments, which may be due to abrasion phenomena at the bone-implant interface. Abrasion phenomena led to a faster insertion of the acetabular cup when the implant had been previously inserted into the same bone cavity. An empirical analytical model considering a flat punch configuration to model the bone-implant contact conditions was used to understand the trend of the variation of I 2 during the insertion of the acetabular cup. The measurement of the force during impacts is useful to assess the bone-implant interface properties, but needs to be validated in the clinic to be useful for orthopaedic surgeons intra-operatively.
用于非骨水泥型髋关节假体植入的压配程序旨在获得最佳的植入物初始稳定性。我们之前曾使用冲击持续时间的测量来跟踪髋臼杯植入物在骨组织中的插入过程。本研究的目的是调查由于髋臼杯连续插入骨组织而导致的冲击动量值的变化。将通过冲击动量和接触持续时间测量获得的结果进行比较。总共对10个牛骨样本进行了三个连续的程序,每个程序包括10次可重复的冲击(3.5千克重物下落40毫米)。每个程序旨在将髋臼杯植入物插入同一个骨腔中。通过力传感器记录每次冲击过程中力随时间的变化,从而测量冲击持续时间(I 1)和动量(I 2)。I 2的值随着冲击次数的增加而增加,并在N 2 = 5.07 ± 1.31次冲击后达到恒定值。此外,统计分析表明,N 2随着实验次数的增加而显著降低,这可能是由于骨-植入物界面的磨损现象所致。当植入物先前已插入同一个骨腔时,磨损现象导致髋臼杯的插入速度更快。使用一个考虑平冲头配置来模拟骨-植入物接触条件的经验分析模型,以了解髋臼杯插入过程中I 2的变化趋势。冲击过程中力的测量有助于评估骨-植入物界面的特性,但需要在临床上进行验证,以便在手术中对骨科医生有用。