Michel Adrien, Bosc Romain, Vayron Romain, Haiat Guillaume
J Biomech Eng. 2015 Mar;137(3). doi: 10.1115/1.4029505. Epub 2015 Feb 2.
The implant primary stability of the acetabular cup (AC) is an important parameter for the surgical success of press-fit procedures used for the insertion of cementless hip prostheses. In previous studies by our group (Mathieu, V., Michel, A., Lachaniette, C. H. F., Poignard, A., Hernigou, P., Allain, J., and Haiat, G., 2013, "Variation of the Impact Duration During the in vitro Insertion of Acetabular Cup Implants," Med. Eng. Phys., 35(11), pp. 1558-1563) and (Michel, A., Bosc, R., Mathieu, V., Hernigou, P., and Haiat, G., 2014, "Monitoring the Press-Fit Insertion of an Acetabular Cup by Impact Measurements: Influence of Bone Abrasion," Proc. Inst. Mech. Eng., Part H, 228(10), pp. 1027-1034), the impact momentum and duration were shown to carry information on the press-fit insertion of the AC within bone tissue. The aim of the present study is to relate the impact momentum recorded during the AC insertion to the AC biomechanical primary stability. The experimental protocol consisted in testing 13 bovine bone samples that underwent successively series of 15 reproducible mass falls impacts (5 kg, 5 cm) followed by tangential stability testing. Each bone sample was tested with different hole sizes in order to obtain different stability configurations. The impact momentum and the tangential primary stability reach a maximum value for an interference fit equal to around 1 mm. Moreover, a correlation between the impact momentum and the stability was obtained with all samples and all configuration (R2 = 0.65). The implant primary stability can be assessed through the measurement of the impact force signal analysis. This study opens new paths for the development of a medical device which could be used as a decision support system to assist the surgeon during the insertion of the AC implant.
髋臼杯(AC)的植入初期稳定性是用于无骨水泥髋关节假体植入的压配手术取得手术成功的一个重要参数。在我们团队之前的研究中(Mathieu, V., Michel, A., Lachaniette, C. H. F., Poignard, A., Hernigou, P., Allain, J., and Haiat, G., 2013, “髋臼杯植入物体外插入过程中冲击持续时间的变化”,《医学工程与物理学》,35(11),第1558 - 1563页)以及(Michel, A., Bosc, R., Mathieu, V., Hernigou, P., and Haiat, G., 2014, “通过冲击测量监测髋臼杯的压配插入:骨磨损的影响”,《机械工程师学会会刊》,H辑,228(10),第1027 - 1034页),冲击动量和持续时间被证明携带着关于AC在骨组织中压配插入的信息。本研究的目的是将AC插入过程中记录的冲击动量与AC的生物力学初期稳定性联系起来。实验方案包括测试13个牛骨样本,这些样本依次接受一系列15次可重复的质量落体冲击(5千克,5厘米),随后进行切向稳定性测试。每个骨样本用不同的孔尺寸进行测试,以获得不同的稳定性配置。对于约1毫米的过盈配合,冲击动量和切向初期稳定性达到最大值。此外,在所有样本和所有配置下都获得了冲击动量与稳定性之间的相关性(R2 = 0.65)。植入初期稳定性可以通过冲击力信号分析的测量来评估。这项研究为一种医疗设备的开发开辟了新途径,该设备可作为决策支持系统,在AC植入物插入过程中辅助外科医生。