眼轴长度对使用光谱域光学相干断层扫描进行厚度测量的影响。

Influence of axial length on thickness measurements using spectral-domain optical coherence tomography.

作者信息

Röck Tobias, Bartz-Schmidt Karl Ulrich, Bramkamp Matthias, Röck Daniel

机构信息

Centre for Ophthalmology, University of Tübingen, Tübingen, Germany.

Department of General Medicine, Ruhr-University Bochum, Bochum, Germany.

出版信息

Invest Ophthalmol Vis Sci. 2014 Oct 8;55(11):7494-8. doi: 10.1167/iovs.14-14043.

Abstract

PURPOSE

The purpose of this study was to assess the influence of axial length on spectral-domain optical coherence tomography (SD-OCT) thickness measurements in patients with subretinal visual implants.

METHODS

Data from eight emmetropic pseudophakic eyes of eight patients with subretinal visual implants were analyzed retrospectively. These patients participated in the monocentric part of a multicenter trial. The axial length was measured in three short (<22.5 mm), three medium (22.51-25.50 mm), and two long (>25.52 mm) eyes. Using Heidelberg Spectralis, the known thickness of a subretinal implant microchip (70 μm) was measured on 15 images per eye with SD-OCT, using the software calipers.

RESULTS

The mean axial length was 20.8 ± 0.8 mm in short eyes, 23.3 ± 0.4 mm in medium eyes, and 26.3 ± 0.5 mm in long eyes. We found in short eyes, in medium eyes, and in long eyes a mean value of microchip thickness measurements from SD-OCT of 82.9 ± 1.4 μm, 70.5 ± 1.3 μm, and 64.2 ± 1.3 μm, respectively. The thickness measurements decreased in SD-OCT measurements with longer axial lengths significantly (P < 0.0001).

CONCLUSIONS

Axial length influences SD-OCT thickness measurements. Our findings demonstrate accuracy of the scaling in SD-OCT thickness measurements in emmetropic medium eyes. Caution is recommended when comparing the measured values of short and long eyes with the normative database of the instrument. There is a need for larger sample-size studies to confirm our results. (ClinicalTrials.gov number, NCT01024803.).

摘要

目的

本研究旨在评估眼轴长度对视网膜下视觉植入患者的频域光学相干断层扫描(SD-OCT)厚度测量的影响。

方法

回顾性分析8例视网膜下视觉植入患者的8只正视人工晶状体眼的数据。这些患者参与了一项多中心试验的单中心部分。测量了3只短眼(<22.5mm)、3只中等眼(22.51 - 25.50mm)和2只长眼(>25.52mm)的眼轴长度。使用海德堡Spectralis,通过软件卡尺在每只眼睛的15张图像上用SD-OCT测量视网膜下植入微芯片的已知厚度(70μm)。

结果

短眼的平均眼轴长度为20.8±0.8mm,中等眼为23.3±0.4mm,长眼为26.3±0.5mm。我们发现,短眼中SD-OCT测量的微芯片厚度平均值为82.9±1.4μm,中等眼中为70.5±1.3μm,长眼中为64.2±1.3μm。随着眼轴长度增加,SD-OCT测量的厚度显著降低(P<0.0001)。

结论

眼轴长度影响SD-OCT厚度测量。我们的研究结果证明了正视中等眼SD-OCT厚度测量中缩放的准确性。在将短眼和长眼的测量值与仪器的标准数据库进行比较时,建议谨慎。需要更大样本量的研究来证实我们的结果。(ClinicalTrials.gov编号,NCT01024803。)