植入可膨胀阴茎假体很久后,出现了管腔侵蚀。

Tubing erosion of an inflatable penile prosthesis long after implantation.

机构信息

Department of Urology, Queen's University Kingston, ON, Canada.

出版信息

Sex Med. 2014 Jun;2(2):103-6. doi: 10.1002/sm2.30.

Abstract

INTRODUCTION

Erosion through skin of connecting tubing of an inflatable penile prosthesis (IPP) has not been previously reported.

AIM

The aim of this study was to present a case of tubing erosion, review the pertinent literature, and discuss the possible causes and management options, including preservation of the device and its components.

METHODS

A 42-year-old male failing to respond to medical treatment for erectile dysfunction underwent insertion of an AMS 700 IPP in 1986. Six years later, a revision was necessary because of a leak in the right cylinder and 4 years after, the pump was replaced. Fourteen years after the original implant, he presented with a portion of the tube connecting the pump to the right cylinder eroding through the skin. There was no infection. The skin area involved was resected and the original pump and tubing were buried in a new scrotal pocket after thorough irrigation.

RESULTS

The IPP remained in place, allowing vaginal penetration and without infection for another 11 years. Three years later, it was de-functionalized, converted into a fixed volume device. It eventually was replaced 25 years after originally implanted with a semirigid prosthesis because it did not provide sufficient rigidity and because of concerns about the presence of "screws" detected during pelvic imaging.

CONCLUSIONS

Mechanical failures in the early IPP models, as illustrated in this case, were expected. However, the long survival of the device is remarkable. Erosion of the connecting tubing through the skin is unique and, under exceptional circumstances, may be managed conservatively without replacing components of the IPP. Clinicians unfamiliar with procedures involving inflatable devices need to be aware of "foreign bodies" visible in radiological examinations in men who have had revisions of an IPP. Morales A. Tubing erosion of an inflatable penile prosthesis long after implantation. Sex Med 2014;2:103-106.

摘要

引言

充气阴茎假体(IPP)连接管穿透皮肤的情况尚未见报道。

目的

本研究旨在报告一例管道侵蚀病例,回顾相关文献,并讨论可能的原因和处理选择,包括设备及其组件的保留。

方法

一位 42 岁男性因勃起功能障碍接受 AMS 700 IPP 植入,1986 年发生故障。6 年后,由于右侧缸体泄漏,需要进行修复;4 年后,更换了泵。在原始植入物 14 年后,他出现了将泵连接到右侧缸体的管段穿透皮肤的情况。无感染。切除受累皮肤区域,在彻底冲洗后,将原始泵和管道埋入新的阴囊袋中。

结果

IPP 仍在位,允许阴道插入,且无感染 11 年。3 年后,它失去功能,转换为固定容量装置。25 年后,由于原始植入物存在“螺丝”,且存在刚性不足问题,更换为半刚性假体。

结论

如本例所示,早期 IPP 模型的机械故障是可以预料的。然而,该设备的长期存活是显著的。连接管穿透皮肤的侵蚀是独特的,在特殊情况下,可以在不更换 IPP 组件的情况下进行保守治疗。不熟悉充气设备操作程序的临床医生需要意识到在进行过 IPP 修复的男性中,放射学检查中可见“异物”。Morales A. 充气阴茎假体植入多年后管道侵蚀。性医学 2014;2:103-106.