Suppr超能文献

Using hand-held dental x-ray devices: ensuring safety for patients and operators.

作者信息

Mahdian Mina, Pakchoian Andrew J, Dagdeviren Didem, Alzahrani Adel, Jalali Elnaz, Tadinada Aditya, Lurie Alan

机构信息

Dr. Mahdian is an oral and maxillofacial radiology resident, Department of Oral Health and Diagnostic Sciences, University of Connecticut School of Dental Medicine, Farmington.

Dr. Pakchoian is an oral and maxillofacial radiology resident, Department of Oral Health and Diagnostic Sciences, University of Connecticut School of Dental Medicine, Farmington.

出版信息

J Am Dent Assoc. 2014 Nov;145(11):1130-2. doi: 10.14219/jada.2014.85.

Abstract

BACKGROUND

Since the introduction of hand-held x-ray units in dentistry, a few inexpensive devices have emerged that lack the necessary safety measures and failed to meet U.S. Food and Drug Administration (FDA) standards. They are advertised actively and sold online in the United States.

METHODS

The authors present several safety issues associated with an imported hand-held x-ray device that has not been cleared by the FDA and compare the device with an FDA-cleared unit.

RESULTS

The authors found that the non-FDA-cleared device posed major safety hazards, including high radiation doses to patients and operators, lack of operator shielding, lower-than-acceptable kilovolt (peak) value, inadequate collimation, lack of an audible signal of x-ray generation and absence of a so-called dead-man switch.

CONCLUSIONS AND PRACTICAL IMPLICATIONS

Dental professionals must be aware of unsafe x-ray equipment and use only those devices that have been cleared by the FDA to protect themselves and their patients.