Kováčik L, Kereïche S, Matula P, Raška I
Institute of Cellular Biology and Pathology, First Faculty of Medicine, Charles University in Prague, Czech Republic.
Centre for Biomedical Image Analysis, Faculty of Informatics, Masaryk University, Brno, Czech Republic.
Folia Biol (Praha). 2014;60 Suppl 1:66-70.
Electron tomographic reconstructions suffer from a number of artefacts arising from effects accompanying the processes of acquisition of a set of tilted projections of the specimen in a transmission electron microscope and from its subsequent computational handling. The most pronounced artefacts usually come from imprecise projection alignment, distortion of specimens during tomogram acquisition and from the presence of a region of missing data in the Fourier space, the "missing wedge". The ray artefacts caused by the presence of the missing wedge can be attenuated by the angular image filter, which attenuates the transition between the data and the missing wedge regions. In this work, we present an analysis of the influence of angular filtering on the resolution of averaged repetitive structural motives extracted from three-dimensional reconstructions of tomograms acquired in the single-axis tilting geometry.
电子断层扫描重建存在许多伪像,这些伪像源于在透射电子显微镜中获取样品的一组倾斜投影过程中伴随的效应,以及随后的计算处理。最明显的伪像通常来自投影对准不精确、断层扫描采集过程中样品的变形以及傅里叶空间中缺失数据区域(即“缺失楔形”)的存在。由缺失楔形导致的射线伪像可以通过角度图像滤波器来减弱,该滤波器会减弱数据区域和缺失楔形区域之间的过渡。在这项工作中,我们分析了角度滤波对从单轴倾斜几何结构中获取的断层扫描三维重建中提取的平均重复结构基元分辨率的影响。