Suppr超能文献

[X射线照片微缩拍摄中的质量保证]

[Quality assurance in the microfilming of x-ray pictures].

作者信息

Haverkamp U, Kronholz H L, Pötter R

机构信息

Klinik und Poliklinik für Strahlentherapie-Radioonkologie, Universität Münster.

出版信息

Rofo. 1989 Apr;150(4):472-4. doi: 10.1055/s-2008-1047059.

Abstract

Microfilming of X-ray films has become more topical owing to the amendment of the X-ray regulations. The storage of X-rays as reproductions on an image or data carrier is expressly envisaged as a possibility in the text of the regulation. The quality of transfer must be insured and the copies must be legible. In microfilming, the latitude and the resolution capacity of the microfilm and remagnification are appropriate as parameters of image quality. Testing of seven filming systems shows a major range of variation of both parameters; in some cases, the results led to reexamination of the camera adjustment and developer chemistry. A test of constancy of filming analogous to direct radiography appears necessary. A corresponding norm does not exist at present.

摘要

由于X射线法规的修订,X射线胶片的微缩拍摄变得更加热门。法规文本明确规定,可将X射线以图像或数据载体复制品的形式进行存储。必须确保传输质量,且副本必须清晰可读。在微缩拍摄中,缩微胶片的宽容度、分辨率及再放大率是图像质量的合适参数。对七种拍摄系统的测试表明,这两个参数的变化范围很大;在某些情况下,测试结果促使人们重新审视相机调整和显影剂化学配方。似乎有必要进行类似于直接放射照相的拍摄稳定性测试。目前尚无相应的标准。