Weber David J, Gergen Maria F, Sickbert-Bennett Emily E, Short Kathleen A, Lanza-Kaduce Kendra E, Rutala William A
Department of Hospital Epidemiology, University of North Carolina (UNC) Health Care, Chapel Hill, North Carolina.
Infect Control Hosp Epidemiol. 2014 Dec;35(12):1543-6. doi: 10.1086/678603. Epub 2014 Nov 10.
Adult hospitalized patients with cystic fibrosis commonly receive nebulized medications. For single-patient-use nebulizers that are cleaned after each use, there is infrequent nebulizer contamination (0%-11%) with only low numbers of epidemiologically important pathogens (less than 100 colony-forming units), and this contamination is similar after 24, 48, and 72 hours of use.
成年住院囊性纤维化患者通常接受雾化药物治疗。对于每次使用后进行清洁的单人使用雾化器,雾化器污染情况不常见(0%-11%),且仅存在数量较少的具有流行病学重要性的病原体(少于100个菌落形成单位),并且在使用24、48和72小时后这种污染情况相似。