Suppr超能文献

旋转血泵搏动性对流入插管尖端血液相容性和血栓形成相关潜在参数的影响。

Effect of rotary blood pump pulsatility on potential parameters of blood compatibility and thrombosis in inflow cannula tips.

作者信息

Wong Kai Chun, Büsen Martin, Benzinger Carrie, Gäng René, Bezema Mirko, Greatrex Nicholas, Schmitz-Rode Thomas, Steinseifer Ulrich

机构信息

Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Helmholtz Institute, RWTH Aachen - Germany.

出版信息

Int J Artif Organs. 2014 Dec;37(12):875-87. doi: 10.5301/ijao.5000361. Epub 2014 Nov 29.

Abstract

PURPOSE

Rotary Blood Pump (RBP) pulsatile strategies relative to the native cardiac cycle have been widely studied because of their benefits to hemodynamics. However, the effects that inducing pulses has on the blood compatibility of ventricular assist device (VAD) support have not yet been understood. Inflow cannulae have been found to be associated with thrombosis under conventional constant speed support of RBPs. To prevent further risks to blood compatibility, it is necessary to understand the relationship between cannula tip design and the induced pulsatility. The purpose of this study was to evaluate the flow field of 5 different tip geometries under RBP pulsatile support using stereo-particle image velocimetry (PIV).

METHODS

Inflow cannulae with conventional tip geometries (blunt, blunt with 4 side ports, beveled with 3 side ports, and cage) and a custom designed crown tip were studied. All cannulae were interposed between a mixed-flow RBP and a silicone left ventricle. The contractile function and hemodynamics were reproduced in a mock circulation loop (MCL). The RBP was configured to induce synchronous and counter-synchronous pulses relative to cardiac cycles while supporting the failing ventricle.

RESULTS

Between both pulsing strategies, low shear volume (γ̇≤100/s, potential parameter of thrombus formation) showed no significant difference. However, counter-synchronous pulsatile mode induced less increase of both high shear volume (γ̇≥2778/s, potential parameter of platelet activation) and recirculation volume (V(z)>0, potential parameter of thrombus formation).

CONCLUSIONS

Although the clinical relationship cannot be inferred from this measurement, when considering the inflow tips only, a necessary trade-off should be made between adverse effects on blood compatibility and benefits for hemodynamics during RBP pulsatile mode.

摘要

目的

由于旋转血泵(RBP)的脉动策略对血流动力学有益,因此已对其与天然心动周期相关的策略进行了广泛研究。然而,诱导脉冲对心室辅助装置(VAD)支持的血液相容性的影响尚未明确。在RBP的传统恒速支持下,已发现流入插管与血栓形成有关。为防止对血液相容性造成进一步风险,有必要了解插管尖端设计与诱导脉动之间的关系。本研究的目的是使用立体粒子图像测速技术(PIV)评估在RBP脉动支持下5种不同尖端几何形状的流场。

方法

研究了具有传统尖端几何形状(钝头、带4个侧孔的钝头、带3个侧孔的斜角和笼式)的流入插管以及一种定制设计的冠状尖端。所有插管均置于混流式RBP和硅胶左心室之间。在模拟循环回路(MCL)中再现收缩功能和血流动力学。RBP被配置为在支持衰竭心室的同时,相对于心动周期诱导同步和反同步脉冲。

结果

在两种脉动策略之间,低剪切体积(γ̇≤100/s,血栓形成的潜在参数)无显著差异。然而,反同步脉动模式导致高剪切体积(γ̇≥2778/s,血小板激活的潜在参数)和再循环体积(V(z)>0,血栓形成的潜在参数)的增加较少。

结论

尽管不能从该测量结果推断临床关系,但仅考虑流入尖端时,在RBP脉动模式期间,应在对血液相容性的不利影响和对血流动力学的益处之间进行必要的权衡。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验