IEEE Trans Med Imaging. 2015 May;34(5):1140-7. doi: 10.1109/TMI.2014.2376813. Epub 2014 Dec 2.
In this work, we demonstrated that an optimized detector angular configuration based on the anisotropic energy distribution of background scattered X-rays improves X-ray fluorescence computed tomography (XFCT) detection sensitivity. We built an XFCT imaging system composed of a bench-top fluoroscopy X-ray source, a CdTe X-ray detector, and a phantom motion stage. We imaged a 6.4-cm-diameter phantom containing different concentrations of gold solution and investigated the effect of detector angular configuration on XFCT image quality. Based on our previous theoretical study, three detector angles were considered. The X-ray fluorescence detector was first placed at 145 (°) (approximating back-scatter) to minimize scatter X-rays. XFCT image quality was compared to images acquired with the detector at 60 (°) (forward-scatter) and 90 (°) (side-scatter). The datasets for the three different detector positions were also combined to approximate an isotropically arranged detector. The sensitivity was optimized with detector in the 145 (°) back-scatter configuration counting the 78-keV gold Kβ1 X-rays. The improvement arose from the reduced energy of scattered X-ray at the 145 (°) position and the large energy separation from gold K β1 X-rays. The lowest detected concentration in this configuration was 2.5 mgAu/mL (or 0.25% Au with SNR = 4.3). This concentration could not be detected with the 60 (°) , 90 (°) , or isotropic configurations (SNRs = 1.3, 0, 2.3, respectively). XFCT imaging dose of 14 mGy was in the range of typical clinical X-ray CT imaging doses. To our knowledge, the sensitivity achieved in this experiment is the highest in any XFCT experiment using an ordinary bench-top X-ray source in a phantom larger than a mouse ( > 3 cm).
在这项工作中,我们证明了基于背景散射 X 射线各向异性能量分布的优化探测器角度配置可以提高 X 射线荧光层析成像(XFCT)的检测灵敏度。我们构建了一个由台式透视 X 射线源、CdTe X 射线探测器和一个仿射运动台组成的 XFCT 成像系统。我们对一个包含不同浓度金溶液的 6.4 厘米直径的仿体进行了成像,并研究了探测器角度配置对 XFCT 图像质量的影响。基于我们之前的理论研究,考虑了三个探测器角度。首先,将 X 射线荧光探测器放置在 145°(近似背散射)以最小化散射 X 射线。将 XFCT 图像质量与探测器放置在 60°(前向散射)和 90°(侧向散射)时获得的图像进行了比较。还将三个不同探测器位置的数据集组合起来以近似各向同性排列的探测器。在 145°背散射配置下优化了探测器的灵敏度,以计数 78keV 的金 Kβ1 X 射线。这种改进源于 145°位置散射 X 射线的能量降低以及与金 Kβ1 X 射线的能量分离较大。在这种配置下,最低检测浓度为 2.5mgAu/mL(或 SNR=4.3 时的 0.25%Au)。60°、90°或各向同性配置(SNR 分别为 1.3、0 和 2.3)无法检测到这个浓度。XFCT 成像剂量为 14 mGy,处于典型临床 X 射线 CT 成像剂量范围内。据我们所知,在使用普通台式 X 射线源在大于老鼠(>3 厘米)的仿体中进行的任何 XFCT 实验中,这个实验达到的灵敏度是最高的。