螺旋断层放疗系统 Hi·Art II 探测器在治疗射线照射下的稳定性。

Stability of the Helical TomoTherapy Hi·Art II detector for treatment beam irradiations.

机构信息

Lausanne University Hospital Institute of Radiation Physics.

出版信息

J Appl Clin Med Phys. 2014 Nov 8;15(6):4897. doi: 10.1120/jacmp.v15i6.4897.

Abstract

The Hi·Art II Helical TomoTherapy (HT) unit is equipped with a built-in onboard MVCT detector used for patient imaging and beam monitoring. Our aim was to study the detector stability for treatment beam measurements. We studied the MVCT detector response with the 6 MV photon beam over time, throughout short-term (during an irradiation) and long-term (two times 50 days) periods. Our results show a coefficient of variation ≤ 1% for detector chambers inside the beam (excluding beam gradients) for short- and long-term response of the MVCT detector. Larger variations were observed in beam gradients and an influence of the X-ray target where degradation was found. The results assume that an 'air scan' procedure is performed daily to recalibrate the detector with the imaging beam. On short term, the detector response stability is comparable to other devices. Long-term measure- ments during two 50-day periods show a good reproducibility.

摘要

Hi·Art II 螺旋断层放射治疗系统(HT)配备了内置的在线 MVCT 探测器,用于患者成像和束流监测。我们的目的是研究治疗束测量中探测器的稳定性。我们研究了 MVCT 探测器在不同时间段内(包括短期照射期间和两个 50 天周期内)对 6MV 光子束的响应。我们的结果表明,对于光束内的探测器室(不包括光束梯度),无论是短期还是长期响应,其变化系数≤1%。在束梯度和 X 射线靶中观察到更大的变化,并且发现存在退化现象。结果假设每天进行“空气扫描”程序,用成像束对探测器进行重新校准。在短期内,探测器的响应稳定性与其他设备相当。在两个 50 天周期的长期测量中,显示出良好的可重复性。