The clinical performance of dental implants is strongly defined by biomechanical principles. The aim of this study was to quantify the Vicker's hardness (VHN) and elastic modulus (E) surround bone to dental implant in different regions, and to discuss the parameters of dynamic microindantion test. Ten cylindrical implants with morse taper interface (Titamax CM, Neodent; 3.5 mm diameter and 7 mm a height) were inserted in rabbit tibia. The mechanical properties were analyzed using microhardness dynamic indenter with 200 mN load and 15 s penetration time. Seven continuous indentations were made distancing 0.08 mm between each other perpendicularly to the implant-bone interface towards the external surface, at the limit of low (Lp) and high implant profile (Hp). Data were analyzed by Student's t-test (a=0.05) to compare the E and VHN values obtained on both regions. Mean and standard deviation of E (GPa) were: Lp. 16.6 ± 1.7, Hp. 17.0 ± 2.5 and VHN (N/mm2): Lp. 12.6 ± 40.8, Hp. 120.1 ± 43.7. No statistical difference was found between bone mechanical properties of high and low profile of the surround bone to implant, demonstrating that the bone characterization homogeneously is pertinent. Dynamic microindantion method proved to be highly useful in the characterization of the individual peri-implant bone tissue.
牙种植体的临床性能在很大程度上由生物力学原理决定。本研究的目的是量化不同区域牙种植体周围骨组织的维氏硬度(VHN)和弹性模量(E),并探讨动态微压痕试验的参数。将10个具有莫氏锥度界面的圆柱形种植体(Titamax CM,Neodent;直径3.5mm,高度7mm)植入兔胫骨。使用加载200mN、穿透时间15s的显微硬度动态压头分析力学性能。在低种植体轮廓(Lp)和高种植体轮廓(Hp)的极限处,垂直于种植体-骨界面,朝着外表面彼此相距0.08mm进行7次连续压痕。通过学生t检验(α=0.05)分析数据,以比较在两个区域获得的E和VHN值。E(GPa)的平均值和标准差为:Lp. 16.6±1.7,Hp. 17.0±2.5;VHN(N/mm2)为:Lp. 12.6±40.8,Hp. 120.1±43.7。种植体周围骨组织高轮廓和低轮廓的骨力学性能之间未发现统计学差异,表明骨特征均匀性是相关的。动态微压痕法被证明在表征单个种植体周围骨组织方面非常有用。