van der Walt T N, Coetzee P P, Fourie P J
Department of Chemistry, Rand Afrikaans University, Johannesburg, Republic of South Africa.
Int J Rad Appl Instrum A. 1989;40(6):525-9. doi: 10.1016/0883-2889(89)90139-1.
N-(4-(n-butyl)-acetanilide)iminodiacetic acid (BIDA), N-(2,6-diisopropylacetanilide)iminodiacetic acid (DISIDA), diethylenetriaminepentaacetic acid (DTPA) and methylene diphosphonic acid (MDP) are used in labelling kits. The contents of BIDA, DISIDA or MDP of the 99mTc-labelled compounds can be determined (indirectly) spectrophotometrically with copper, eriochrome cyaanine R (ECC) and dodecylethyldimethylammonium bromide (DEDA) in a sodium barbital buffered system at pH 8.5. The calibration curves obey Beer's Law from 0 to 40 micrograms/25 mL for BIDA and DISIDA, 0 to 60 micrograms/25 mL for DTPA and 0 to 100 micrograms/10 mL for MDP.
N-(4-(正丁基)-乙酰苯胺基)亚氨基二乙酸(BIDA)、N-(2,6-二异丙基乙酰苯胺基)亚氨基二乙酸(DISIDA)、二亚乙基三胺五乙酸(DTPA)和亚甲基二膦酸(MDP)用于标记试剂盒。在pH 8.5的巴比妥钠缓冲体系中,99mTc标记化合物中的BIDA、DISIDA或MDP含量可以用铜、铬菁R(ECC)和十二烷基乙基二甲基溴化铵(DEDA)通过分光光度法(间接)测定。BIDA和DISIDA的校准曲线在0至40微克/25毫升范围内符合比尔定律,DTPA在0至60微克/25毫升范围内,MDP在0至100微克/10毫升范围内。