非创伤性种植体取出:绵羊胫骨的生物力学和生物学分析

Nontraumatic Implant Explantation: A Biomechanical and Biological Analysis in Sheep Tibia.

作者信息

Anitua Eduardo, Murias-Freijo Alia, Piñas Laura, Tejero Ricardo, Prado Roberto, Orive Gorka

机构信息

1  Private practice in implantology and oral rehabilitation, Vitoria, Spain.

2  Biotechnology Institute (BTI), Vitoria, Spain.

出版信息

J Oral Implantol. 2016 Feb;42(1):3-11. doi: 10.1563/aaid-joi-D-14-00193. Epub 2014 Dec 23.

Abstract

Preclinical research in a sheep tibia model has been conducted to evaluate the underlying mechanisms of the nontraumatic implant explantation of failed implants, which allow placing a new one in the bone bed. Twelve dental implants were placed in sheep diaphysis tibia and once osseointegrated they were explanted using a nontraumatic implant explantation approach. Implant osseointegration and explantation were monitored by means of frequency resonance, removal torque, and angle of rotation measurement. The host bone bed and the explanted implant surface were analyzed by conventional microscopy and scanning electron microscope. Results show that osseointegration was broken with an angular displacement of less than 20°. In this situation the implant returns to implant stability quotient values in the same range of their primary stability. Moreover, the explantation technique causes minimal damage to the surrounding bone structure and cellularity. This nontraumatic approach allows the straightforward replacement of failed implants and emerges as a promising strategy to resolve clinically challenging situations.

摘要

在绵羊胫骨模型中进行了临床前研究,以评估失败种植体非创伤性取出的潜在机制,这使得能够在骨床中植入新的种植体。将12颗牙种植体植入绵羊胫骨骨干,一旦骨整合完成,就采用非创伤性种植体取出方法将其取出。通过频率共振、去除扭矩和旋转角度测量来监测种植体的骨整合和取出情况。通过传统显微镜和扫描电子显微镜对宿主骨床和取出的种植体表面进行分析。结果表明,骨整合在角位移小于20°时被破坏。在这种情况下,种植体的种植体稳定性商值恢复到其初始稳定性的相同范围内。此外,取出技术对周围骨结构和细胞的损伤最小。这种非创伤性方法允许直接替换失败的种植体,并成为解决临床挑战性情况的一种有前景的策略。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验