使用IOLMaster设备在6个月的时间间隔内对初始眼轴长度、角膜曲率读数和人工晶状体度数进行个体内比较。

Intrapersonal comparison of initial axial length, keratometric readings, and intraocular lens power over a 6-month interval using an IOLMaster device.

作者信息

Khambhiphant Bharkbhum, Varadisai Adisai, Visalvate Nat

机构信息

Department of Ophthalmology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Chulalongkorn University, Bangkok, Thailand.

Department of Ophthalmology, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.

出版信息

Clin Ophthalmol. 2014 Dec 22;9:21-4. doi: 10.2147/OPTH.S75075. eCollection 2015.

Abstract

PURPOSE

In order to improve current biometry practice, we investigated changes in axial length (AL), keratometry (K), and intraocular lens (IOL) power measured by the IOLMaster in adults within a 6-month period at the outpatient eye clinic, King Chulalongkorn Memorial Hospital, Bangkok, Thailand.

PATIENTS AND METHODS

This was a nonrandomized prospective cohort study. The AL, steepest K, flattest K, and IOL power of nondilated eyes were measured at the initial examination and later at 6 months by the IOLMaster. Changes in AL, K, and IOL power using the SRK (Sanders, Retzlaff, and Kraff)-T formula within 6 months were compared using the paired-sample t-test.

RESULTS

The study comprised 90 eyes from 90 patients. The mean age of the patients was 62.99±13.95 years (range 28-87 years). Mean values and standard deviation for AL, steepest K, flattest K, and IOL power at initial visit and 6 months later were 23.56±1.51 mm and 23.56±1.52 mm (P=1.0; 95% confidence interval [CI] -0.01 to 0), 44.94±1.37 D and 44.98±1.43 D (P=0.96; 95% CI -0.13 to 0.04), 43.85±1.43 D and 43.89±1.36 D (P=0.93; 95% CI -0.02 to 0.07), and 20.27±4.12 D and 20.15±4.51 D (P=0.99; 95% CI -0.05 to 0.29), respectively.

CONCLUSION

There was no statistically significant difference in AL (P=1.0), both K (P=0.96, P=0.93), or IOL power (P=0.99) within the 6-month period using the IOLMaster. Remeasurement at 6 months may be required in some cases, but should not be routine.

摘要

目的

为改进当前的生物测量实践,我们在泰国曼谷朱拉隆功国王纪念医院门诊眼科诊所,对成年人在6个月内使用IOLMaster测量的眼轴长度(AL)、角膜曲率(K)和人工晶状体(IOL)度数的变化进行了调查。

患者与方法

这是一项非随机前瞻性队列研究。在初次检查时以及6个月后,使用IOLMaster测量未散瞳眼睛的AL、最陡K值、最平K值和IOL度数。使用配对样本t检验比较6个月内使用SRK(桑德斯、雷茨拉夫和克拉夫)-T公式计算的AL、K和IOL度数的变化。

结果

该研究纳入了90例患者的90只眼睛。患者的平均年龄为62.99±13.95岁(范围28 - 87岁)。初次就诊时和6个月后的AL、最陡K值、最平K值和IOL度数的平均值及标准差分别为23.56±1.51mm和23.56±1.52mm(P = 1.0;95%置信区间[CI] - 0.01至0)、44.94±1.37D和44.98±1.43D(P = 0.96;95%CI - 0.13至0.04)、43.85±1.43D和43.89±1.36D(P = 0.93;95%CI - 0.02至0.07)以及20.27±4.12D和20.15±4.51D(P = 0.99;95%CI - 0.05至0.29)。

结论

使用IOLMaster在6个月内,AL(P = 1.0)、K值(P = 0.96,P = 0.93)或IOL度数(P = 0.99)均无统计学上的显著差异。在某些情况下可能需要在6个月时重新测量,但不应作为常规操作。

相似文献

引用本文的文献

本文引用的文献

9
Global data on visual impairment in the year 2002.2002年全球视力损害数据。
Bull World Health Organ. 2004 Nov;82(11):844-51. Epub 2004 Dec 14.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验