The aim of this study was to assess the impact of cataract progression using the Haigis formula-calculated intraocular lens (IOL) power and investigate the accuracy of IOL power measured at different time points.
This prospective study was performed on 75 eyes of 75 patients who underwent uneventful cataract surgery. Preoperative ocular parameters including axial length (AL), keratometry (K), anterior chamber depth (ACD), corneal astigmatism, corrected distance visual acuity (CDVA) and uncorrected distance visual acuity (UDVA) examined at the two time points, more than 3 months preoperatively and preoperative 1 day were compared. The ocular parameters measured in the two time points were used to calculate the predicted implanted IOL power and the actual IOL power was chosen on the basis of parameters measured earlier before surgery using the Haigis formula. The mean numerical error (MNE) and mean absolute error (MAE) predicted by the two time points were also compared.
There were significant differences in the ACD, IOL power, UDVA and CDVA (P<0.01), but no statistical differences in AL, mean K and corneal astigmatism (P>0.05) during the average of 5.6 months before surgery. No statistically significant difference was detected in MNE (P>0.05), while the MAE had a significant difference in the two time points (P<0.05).
The IOL power measured earlier before surgery might result in a higher accuracy and the postoperative refractive outcome tended towards emmetropia.
本研究旨在评估使用 Haigis 公式计算的人工晶状体(IOL)屈光力对白内障进展的影响,并研究不同时间点测量的 IOL 屈光力的准确性。
本前瞻性研究纳入 75 例 75 只眼,这些眼均行无并发症的白内障手术。比较术前 3 个月以上和术前 1 天的 2 个时间点的眼轴长度(AL)、角膜曲率(K)、前房深度(ACD)、角膜散光、矫正远视力(CDVA)和未矫正远视力(UDVA)等眼部参数。使用 2 个时间点测量的眼部参数计算预测植入 IOL 屈光力,并根据手术前较早时间点测量的参数使用 Haigis 公式选择实际 IOL 屈光力。还比较了 2 个时间点预测的平均数值误差(MNE)和平均绝对误差(MAE)。
在术前平均 5.6 个月期间,ACD、IOL 屈光力、UDVA 和 CDVA 存在显著差异(P<0.01),但 AL、平均 K 和角膜散光无统计学差异(P>0.05)。MNE 无统计学差异(P>0.05),而 MAE 在 2 个时间点有显著差异(P<0.05)。
手术前较早时间点测量的 IOL 屈光力可能具有更高的准确性,术后屈光状态趋于正视。