Suppr超能文献

利用伦敦植入物取出中心的数据,对英格兰、威尔士和北爱尔兰国家关节注册中心的初次金属对金属髋关节置换术进行验证:一项使用国家关节注册中心数据集的研究

Validation of primary metal-on-metal hip arthroplasties on the National Joint Registry for England, Wales and Northern Ireland using data from the London Implant Retrieval Centre: a study using the NJR dataset.

作者信息

Sabah S A, Henckel J, Cook E, Whittaker R, Hothi H, Pappas Y, Blunn G, Skinner J A, Hart A J

机构信息

Royal National Orthopaedic Hospital, Brockley Hill, Stanmore, HA7 4LP, UK.

Hillingdon Hospital, Uxbridge, London, UK.

出版信息

Bone Joint J. 2015 Jan;97-B(1):10-8. doi: 10.1302/0301-620X.97B1.35279.

Abstract

Arthroplasty registries are important for the surveillance of joint replacements and the evaluation of outcome. Independent validation of registry data ensures high quality. The ability for orthopaedic implant retrieval centres to validate registry data is not known. We analysed data from the National Joint Registry for England, Wales and Northern Ireland (NJR) for primary metal-on-metal hip arthroplasties performed between 2003 and 2013. Records were linked to the London Implant Retrieval Centre (RC) for validation. A total of 67,045 procedures on the NJR and 782 revised pairs of components from the RC were included. We were able to link 476 procedures (60.9%) recorded with the RC to the NJR successfully. However, 306 procedures (39.1%) could not be linked. The outcome recorded by the NJR (as either revised, unrevised or death) for a primary procedure was incorrect in 79 linked cases (16.6%). The rate of registry-retrieval linkage and correct assignment of outcome code improved over time. The rates of error for component reference numbers on the NJR were as follows: femoral head category number 14/229 (5.0%); femoral head batch number 13/232 (5.3%); acetabular component category number 2/293 (0.7%) and acetabular component batch number 24/347 (6.5%). Registry-retrieval linkage provided a novel means for the validation of data, particularly for component fields. This study suggests that NJR reports may underestimate rates of revision for many types of metal-on-metal hip replacement. This is topical given the increasing scope for NJR data. We recommend a system for continuous independent evaluation of the quality and validity of NJR data.

摘要

关节置换登记系统对于关节置换手术的监测及结果评估至关重要。登记数据的独立验证可确保其高质量。目前尚不清楚骨科植入物取出中心验证登记数据的能力如何。我们分析了英格兰、威尔士和北爱尔兰国家关节登记处(NJR)2003年至2013年期间进行的初次金属对金属髋关节置换手术的数据。记录与伦敦植入物取出中心(RC)的数据相链接以进行验证。NJR记录了总共67045例手术,RC有782对经修订的组件。我们成功地将RC记录的476例手术(60.9%)与NJR相链接。然而,306例手术(39.1%)无法链接。在79例已链接的病例中(16.6%),NJR记录的初次手术结果(修订、未修订或死亡)有误。登记处与取出中心的链接率以及结果代码的正确分配率随时间有所提高。NJR上组件参考编号的错误率如下:股骨头类别编号14/229(5.0%);股骨头批次编号13/232(5.3%);髋臼组件类别编号2/293(0.7%);髋臼组件批次编号24/347(6.5%)。登记处与取出中心的链接为数据验证提供了一种新方法,特别是对于组件字段。本研究表明,NJR报告可能低估了许多类型金属对金属髋关节置换的翻修率。鉴于NJR数据范围不断扩大,这一问题备受关注。我们建议建立一个对NJR数据的质量和有效性进行持续独立评估的系统。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4e1d/4548488/1a105971f185/35279-galleyfig1.jpg