Itty Sujit, Callanan David, Jones Racheal, Pecen Paula, Martel Joseph, Jaffe Glenn J
Department of Ophthalmology, Duke University Medical Center, Durham, North Carolina.
Texas Retina Associates, Arlington, Texas.
Am J Ophthalmol. 2015 May;159(5):868-76.e1. doi: 10.1016/j.ajo.2015.01.031. Epub 2015 Jan 30.
To characterize spontaneous dislocation of the fluocinolone acetonide sustained release implant (Retisert; Bausch & Lomb) drug pellet from its strut.
Retrospective, observational case series.
The medical records of 224 patients in whom 407 fluocinolone acetonide implants were placed by either of 2 surgeons between 1998 and 2012 were reviewed for drug pellet-suture strut separation, observed before surgery in clinic or occurring at the time of implant replacement.
Seventeen (4.2%) of 407 drug pellets spontaneously dislocated away from the suture strut at a mean postimplantation time of 77.4 months (range, 33 to 132 months). The rate of spontaneous dislocation increased from 1 dislocation per 262 person years (95% confidence interval, 1 per 1269 to 1 per 89.6 person-years) before 60 months to 1 dislocation per 13.3 person-years (95% confidence interval, 1 per 24.4 to 1 per 7.9 person-years) afterward. Fourteen of 17 dislocated pellets were removed surgically. Of 77 exchanged fluocinolone acetonide implants without dislocated pellets, 11 (14.2%) pellets had separated, but not dislocated away, from the strut; time to exchange of dislocated or dissociated pellets was significantly longer than time to exchange of intact fluocinolone acetonide implants (70.5 vs 44.9 months; P < .001).
The fluocinolone acetonide drug pellet can dislocate spontaneously from the suture strut as a late event that typically occurs after 3 years. Pellets also may be dissociated from the strut without dislocation, when separation occurs at the time of surgical replacement. Removal of the dislocated pellet can be achieved safely. Time was the major risk factor identified for both events.
描述醋酸氟轻松缓释植入物(Retisert;博士伦公司)药物小球从其支柱自发脱位的情况。
回顾性观察病例系列。
回顾了1998年至2012年间由2位外科医生之一为224例患者植入407枚醋酸氟轻松植入物的病历,以检查药物小球与缝线支柱的分离情况,这种分离在术前临床检查时观察到或在植入物更换时出现。
407枚药物小球中有17枚(4.2%)在植入后平均77.4个月(范围33至132个月)时自发从缝线支柱脱位。自发脱位率从60个月前每262人年1例脱位(95%置信区间,每1269人年1例至每89.6人年1例)增加到之后每13.3人年1例脱位(95%置信区间,每24.4人年1例至每7.9人年1例)。17枚脱位小球中有14枚通过手术取出。在77枚更换的未脱位醋酸氟轻松植入物中,有11枚(14.2%)小球与支柱分离但未脱位;脱位或分离小球的更换时间显著长于完整醋酸氟轻松植入物的更换时间(70.5个月对44.9个月;P <.001)。
醋酸氟轻松药物小球可作为一种晚期事件自发从缝线支柱脱位,通常发生在3年后。当在手术更换时发生分离时,小球也可能与支柱解离但未脱位。脱位小球可安全取出。时间是这两种情况确定的主要危险因素。