两种可生物吸收药物洗脱冠状动脉支架(Absorb 和 DESolve)与一种持久金属药物洗脱支架(ML8/Xpedition)的独立体外对比研究。

An independent bench comparison of two bioresorbable drug-eluting coronary scaffolds (Absorb and DESolve) with a durable metallic drug-eluting stent (ML8/Xpedition).

机构信息

Mercy Angiography, Auckland, New Zealand.

出版信息

EuroIntervention. 2015 May;11(1):60-7. doi: 10.4244/EIJY15M02_03.

Abstract

AIMS

We compared the mechanical and physical properties and the safety from strut fracture of side branch and post-dilatation strategies for the Absorb and DESolve bioresorbable scaffolds with the durable metallic drug-eluting XIENCE Xpedition stent using largely independent bench testing.

METHODS AND RESULTS

The strut thickness and crossing profile of the polymeric scaffolds was greater than those of the metallic drug-eluting stent. While all three devices recoiled after deployment, the DESolve enlarged between 10 mins and one hour returning to the immediate post-deployment diameter ("self-correction"). In 3.0 mm stents/scaffolds, the main branch post-dilatation safe threshold without fracture for Absorb was 3.8 mm at 20 atm, for DESolve was 5.0 mm at 20 atm whereas the ML8 did not fracture. For side branch dilatation with a 3.0 mm non-compliant balloon, the threshold before the Absorb fractured was 10 atm whereas the DESolve and ML8 did not fracture at 22 atm. The safe threshold for mini-kissing balloon post-dilatation in 3.0 mm scaffolds/stents with 3.0 mm non-compliant balloons was 5 atm for the Absorb whereas the DESolve and ML8 did not fracture up to 20 atm.

CONCLUSIONS

The metallic stent has thinner struts, lower profile, and greater radial strength than the polymeric scaffolds. Different safe pressure thresholds exist for different scaffolds/stents. Unlike the others, the DESolve showed "self-correction" or enlargement after initial recoil.

摘要

目的

我们使用独立的台架试验比较了 Absorb 和 DESolve 生物可吸收支架与耐用型金属药物洗脱 XIENCE Xpedition 支架的机械和物理性能以及分支支架断裂的安全性。

方法和结果

聚合物支架的支架厚度和交叉轮廓大于金属药物洗脱支架。虽然所有三种装置在部署后都回缩,但 DESolve 在 10 分钟至 1 小时之间扩大,恢复到即时部署后的直径(“自我纠正”)。在 3.0 毫米支架/支架中,Absorb 主支后扩张无断裂的安全阈值为 3.8 毫米,20 个大气压;DESolve 的安全阈值为 5.0 毫米,20 个大气压;而 ML8 没有断裂。对于 3.0 毫米非顺应性球囊的分支扩张,Absorb 断裂前的阈值为 10 个大气压,而 DESolve 和 ML8 在 22 个大气压下没有断裂。对于 3.0 毫米支架/支架的 3.0 毫米非顺应性球囊的迷你吻球囊后扩张,Absorb 的安全阈值为 5 个大气压,而 DESolve 和 ML8 直到 20 个大气压都没有断裂。

结论

金属支架的支架更薄,轮廓更低,径向强度更大,比聚合物支架。不同的支架/支架有不同的安全压力阈值。与其他支架不同,DESolve 在初始回缩后显示出“自我纠正”或扩张。