具有无创电生理重新测试能力的外部激活抗心动过速起搏器。

Externally activated antitachycardia pacemaker with noninvasive electrophysiologic re-testing capability.

作者信息

Aonuma K, Rozanski J J, Barold S S, DeWitt P L, Gosselin A J, Lister J W

出版信息

Pacing Clin Electrophysiol. 1985 Mar;8(2):215-24. doi: 10.1111/j.1540-8159.1985.tb05752.x.

Abstract

An Omni Orthocor 234A special device pacemaker was implanted in nine patients for treatment of drug-resistant supraventricular tachycardia (three patients) or ventricular tachycardia (six patients). This device is activated using a special external unit, which delivers from six to 15 stimuli at preselected coupling intervals ranging from 195 to 500 ms. Serial noninvasive electrophysiology studies were performed either in the hospital or on an outpatient basis using the triggered mode, which can respond to programmed chest wall stimulation. Three patients with reciprocating supraventricular tachycardia were treated effectively by this device during an average follow-up of 2.2 years. Of the six patients with stable ventricular tachycardia amenable to pacemaker termination, only four remained successful during an average follow-up of 1.9 years. Such devices may find important use in serial noninvasive electrophysiologic testing during long-term clinical follow-up. The present form of this device is inadequate for widespread application, but may be useful for highly selected individuals.

摘要

9名患者植入了一台Omni Orthocor 234A特殊装置起搏器,用于治疗耐药性室上性心动过速(3例)或室性心动过速(6例)。该装置通过一个特殊的外部装置激活,该外部装置以195至500毫秒的预选耦合间期发送6至15次刺激。在医院或门诊进行连续无创电生理研究,采用触发模式,该模式可对程控胸壁刺激作出反应。3例折返性室上性心动过速患者在平均2.2年的随访期间通过该装置得到有效治疗。在6例适合起搏器终止的稳定性室性心动过速患者中,平均1.9年的随访期间只有4例仍成功。此类装置在长期临床随访的连续无创电生理检测中可能会有重要用途。该装置的当前形式不足以广泛应用,但可能对经过严格挑选的个体有用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验