Kersschot I E, Ortmanns P, Goethals M A
Pacing Clin Electrophysiol. 1985 May;8(3 Pt 1):402-7. doi: 10.1111/j.1540-8159.1985.tb05779.x.
Atrial pacing bigeminy, defined as atrial pacing with alternating short and long cycle lengths, was recorded in two out of five DDD-pacemaker implants. Another fascinating feature, the lowering of the effective pacing rate after programming to a higher rate, occurred in one patient. Both phenomena can be explained by crosstalk; appropriate programming resulted in normal DDD-function in all cases. These observations in our patients have led to some interesting conclusions regarding multiprogrammability and nominal settings.
在五例DDD起搏器植入患者中,有两例记录到心房起搏二联律,其定义为心房起搏时周期长度长短交替。在一名患者中出现了另一个有趣的特征,即程控为较高频率后有效起搏频率降低。这两种现象都可以用串扰来解释;所有病例中通过适当的程控均实现了正常的DDD功能。我们对这些患者的观察得出了一些关于多程控性和标称设置的有趣结论。