J Cataract Refract Surg. 2014 Feb;40(2):269-75. doi: 10.1016/j.jcrs.2013.07.046. Epub 2013 Dec 22.
To evaluate the repeatability of the anterior and posterior corneal wavefront aberrations using the Sirius Scheimpflug-Placido topographer in normal eyes and keratoconus eyes.
Bozok University Faculty of Medicine, Yozgat, Turkey.
Evaluation of diagnostic test.
In eyes of healthy subjects and eyes of keratoconus patients, 3 repeated measurements were obtained using the Scheimpflug-Placido topographer. Repeatability of the corneal aberrometric data using a 7th-order Zernike expansion (6.0 mm pupil) and central corneal power (3.0 mm zone) in the anterior and posterior corneal surfaces were analyzed. The within-subject standard deviation (Sw) and the intraclass correlation coefficient (ICC) were calculated.
For all modal pairs, the Sw was 0.08 μm or less for anterior and posterior corneal aberrations in both groups. The ICC of the anterior corneal surface ranged from 0.607 (pentafoil) to 0.988 (primary coma) in keratoconus eyes (n = 41) and from 0.568 (quadrifoil) to 0.856 (primary coma) in normal eyes (n = 30). The ICCs for posterior corneal surface aberrometry were 0.656 to 0.873 and 0.592 to 0.824, respectively. For anterior and posterior corneal curvatures, the Sw was 0.12 or lower and the ICC values were more than 0.93 in all cases except the posterior corneal surface reading at the 3.0 mm corneal area in keratoconus eyes (ICC 0.875).
The intraexaminer repeatability of most anterior corneal aberrations with the Scheimpflug-Placido system was moderate to high in normal eyes and keratoconus eyes. The system showed moderate repeatability for the posterior corneal surface.
No author has a financial or proprietary interest in any material or method mentioned.
使用 Sirius Scheimpflug-Placido topography 评估正常眼和圆锥角膜眼中的前、后角膜波前像差的可重复性。
土耳其约兹加特博佐克大学医学院。
诊断测试评估。
在健康受试者的眼中和圆锥角膜患者的眼中,使用 Scheimpflug-Placido topography 获得了 3 次重复测量。使用第 7 阶 Zernike 展开(6.0mm 瞳孔)和前、后角膜表面的中央角膜屈光力(3.0mm 区)分析角膜像差数据的可重复性。计算个体内标准差(Sw)和组内相关系数(ICC)。
对于所有模态对,在两组中,前、后角膜像差的 Sw 均为 0.08μm 或更小。圆锥角膜眼(n=41)前角膜表面的 ICC 范围为 0.607(五叶形)至 0.988(初级彗差),正常眼(n=30)为 0.568(四叶形)至 0.856(初级彗差)。后角膜表面像差的 ICC 值为 0.656 至 0.873 和 0.592 至 0.824。对于前、后角膜曲率,除了圆锥角膜眼的 3.0mm 角膜区域的后角膜表面读数(ICC 0.875)外,Sw 为 0.12 或更低,ICC 值均大于 0.93。
在正常眼和圆锥角膜眼中,Scheimpflug-Placido 系统的大多数前角膜像差的观察者内可重复性为中度至高度。该系统对后角膜表面具有中等的可重复性。