To evaluate the impact of reused phaco tip on intraoperative performance and postoperative outcomes after 2.2 mm micro-coaxial torsional and transversal phacoemulsification.
This prospective randomized study enrolled 136 eyes of 136 cataract patients; 68 eyes in torsional group (34 eyes with single use tip, 34 eyes with reused tip) and 68 eyes in transversal group (34 eyes with single use tip, 34 eyes with reused tip). Intraoperative measurements were total ultrasound (U/S) time, torsional U/S time, cumulative dissipated energy (CDE), estimated fluid use (EFU) in the torsional group and total phacoemulsification time (TPT), and effective phacoemulsification time (EPT) in transversal phacoemulsification. The central endothelial cell density (ECD) and the central corneal thickness (CCT) were evaluated preoperatively and postoperatively at 1 and 30 days using noncontact specular microscopy.
Intraoperative measurements in torsional phacoemulsification showed significant increase in total U/S time (p = 0.01), torsional U/S time (p = 0.01), and CDE (p = 0.01) with the reused tip. The EFU was similar in both tip groups (p = 0.36). The total U/S time, torsional U/S time, and CDE in torsional group were significantly increased with the reused tip for grade III (p = 0.03 for all parameters) and grade IV cataracts (p = 0.005 for torsional U/S time and CDE; p = 0.006 for total U/S time). Intraoperative measurements in transversal phacoemulsification showed no differences in TPT and EPT between tip groups (p > 0.05). The change in ECD and CCT in torsional and transversal phacoemulsification groups were similar in both tip groups (p > 0.05).
This study showed that sharpness of phaco tip edge increases the efficiency of torsional phacoemulsification for hard cataracts. This result does not mean that reused phaco tip may be used in soft and medium cataracts. As the timing of tip's change is a user-dependent procedure, a new single phaco tip should be used in each case if possible.
评估重复使用的超声乳化针头对2.2毫米微同轴扭转和横向超声乳化术中操作及术后结果的影响。
这项前瞻性随机研究纳入了136例白内障患者的136只眼;扭转组68只眼(34只眼使用一次性针头,34只眼使用重复使用的针头),横向组68只眼(34只眼使用一次性针头,34只眼使用重复使用的针头)。术中测量指标为扭转组的总超声(U/S)时间、扭转U/S时间、累积消散能量(CDE)、估计液体用量(EFU),以及横向超声乳化术中的总超声乳化时间(TPT)和有效超声乳化时间(EPT)。使用非接触式镜面显微镜在术前以及术后1天和30天评估中央内皮细胞密度(ECD)和中央角膜厚度(CCT)。
扭转超声乳化术中的术中测量显示,使用重复使用的针头时,总U/S时间(p = 0.01)、扭转U/S时间(p = 0.01)和CDE(p = 0.01)显著增加。两组针头的EFU相似(p = 0.36)。对于III级(所有参数p = 0.03)和IV级白内障(扭转U/S时间和CDE p = 0.005;总U/S时间p = 0.006),扭转组使用重复使用的针头时,总U/S时间、扭转U/S时间和CDE显著增加。横向超声乳化术中的术中测量显示,两组针头的TPT和EPT无差异(p > 0.05)。扭转和横向超声乳化组中,两组针头的ECD和CCT变化相似(p > 0.05)。
本研究表明,超声乳化针头边缘的锋利度可提高硬性白内障扭转超声乳化的效率。这一结果并不意味着重复使用的超声乳化针头可用于软性和中等程度的白内障。由于更换针头的时机取决于操作者,因此在每种情况下尽可能使用新的单个超声乳化针头。