Suppr超能文献

起搏器电极的自发移位及其随后的成功重新植入。

Spontaneous displacement of a pacemaker electrode and its subsequent successful reimplantation.

作者信息

Gialafos J, Theophilis A, Siamas G

机构信息

B Department of Cardiology, University of Athens Medical School, Hippokration Hospital, Greece.

出版信息

Br Heart J. 1989 Nov;62(5):403-5. doi: 10.1136/hrt.62.5.403.

Abstract

A case of what was assumed to be twiddler's syndrome was caused by spontaneous twisting and displacement of the electrode on the fortieth postoperative day. The pacemaker had not been manipulated by the patient, the pocket was tight, and the proximal electrode was well secured. The complication was managed successfully by reimplanting the same electrode after stiffening the lead near the generator with a portion of the stylet.

摘要

一例疑似旋弄综合征的病例是由术后第40天电极自发扭转和移位引起的。患者未对起搏器进行操作,起搏器囊袋紧致,近端电极固定良好。通过用一部分探条在发生器附近加固导线后重新植入同一电极,成功处理了该并发症。