Suppr超能文献

洛伦兹力电阻抗断层成像中的声电散斑图案

Acousto-electrical speckle pattern in Lorentz force electrical impedance tomography.

作者信息

Grasland-Mongrain Pol, Destrempes François, Mari Jean-Martial, Souchon Rémi, Catheline Stefan, Chapelon Jean-Yves, Lafon Cyril, Cloutier Guy

机构信息

Laboratoire de Biorhéologie et d'Ultrasonographie Médicale, Centre Hospitalier de l'Université de Montréal (CRCHUM), Tour Viger, 900 rue Saint-Denis, Montréal (QC) H2X 0A9, Canada.

出版信息

Phys Med Biol. 2015 May 7;60(9):3747-57. doi: 10.1088/0031-9155/60/9/3747. Epub 2015 Apr 23.

Abstract

Ultrasound speckle is a granular texture pattern appearing in ultrasound imaging. It can be used to distinguish tissues and identify pathologies. Lorentz force electrical impedance tomography is an ultrasound-based medical imaging technique of the tissue electrical conductivity. It is based on the application of an ultrasound wave in a medium placed in a magnetic field and on the measurement of the induced electric current due to Lorentz force. Similarly to ultrasound imaging, we hypothesized that a speckle could be observed with Lorentz force electrical impedance tomography imaging. In this study, we first assessed the theoretical similarity between the measured signals in Lorentz force electrical impedance tomography and in ultrasound imaging modalities. We then compared experimentally the signal measured in both methods using an acoustic and electrical impedance interface. Finally, a bovine muscle sample was imaged using the two methods. Similar speckle patterns were observed. This indicates the existence of an 'acousto-electrical speckle' in the Lorentz force electrical impedance tomography with spatial characteristics driven by the acoustic parameters but due to electrical impedance inhomogeneities instead of acoustic ones as is the case of ultrasound imaging.

摘要

超声散斑是超声成像中出现的一种颗粒状纹理图案。它可用于区分组织和识别病变。洛伦兹力电阻抗断层成像技术是一种基于超声的组织电导率医学成像技术。它基于在置于磁场中的介质中施加超声波,并测量由洛伦兹力感应产生的电流。与超声成像类似,我们推测在洛伦兹力电阻抗断层成像中也能观察到散斑。在本研究中,我们首先评估了洛伦兹力电阻抗断层成像与超声成像模态中测量信号之间的理论相似性。然后,我们使用声阻抗和电阻抗界面通过实验比较了两种方法测量的信号。最后,使用这两种方法对牛肌肉样本进行成像。观察到了相似的散斑图案。这表明在洛伦兹力电阻抗断层成像中存在“声电散斑”,其空间特征由声学参数驱动,但归因于电阻抗不均匀性,而非像超声成像那样归因于声学不均匀性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验