柄部锥度再利用对陶瓷股骨头失效载荷的影响。

The influence of stem taper re-use upon the failure load of ceramic heads.

作者信息

Gührs Julian, Krull Annika, Witt Florian, Morlock Michael M

机构信息

TUHH Hamburg University of Technology, Denickestrasse 15, 21073 Hamburg, Germany.

出版信息

Med Eng Phys. 2015 Jun;37(6):545-52. doi: 10.1016/j.medengphy.2015.03.010. Epub 2015 Apr 20.

Abstract

Ceramic components are frequently used in total hip replacement due to their good tribological properties. In revision of broken ceramic heads clinical uncertainties arise, whether the taper of the stem can be re-used with a new ceramic head, especially if the stem is well fixed. Ceramic is a brittle material. Even small damage on the male stem taper can lead to stress concentrations causing premature failure of a new ceramic head. As a consequence, manufactures strictly prohibit stem taper re-use for ceramic heads. The aim of this study was to determine the fracture strength of ceramic heads assembled to re-used male stem tapers, which were subjected to prior head fracture. Five 12/14 Ti6Al4V male tapers and 15 Al2O3 ceramic heads (BIOLOX forte(®); ∅ 28 mm, L) were used for three consecutive fracture tests. Before and after every fracture test, all components were inspected visually and the surface geometry was analyzed. Mean fracture force (52.5 kN) did not decrease with the number of taper re-uses (p ≥ 0.77) but the range increased significantly from initially 4.1 kN to 31.8 kN for the first and 52.6 kN for the second re-use due to some components failing at very low loads. Visual inspection was not sufficient to predict the reduced failure loads. Ceramic heads should therefore not be put on used male tapers without metal adapter sleeves.

摘要

由于陶瓷部件具有良好的摩擦学性能,因此常用于全髋关节置换。在翻修破损的陶瓷股骨头时,临床上面临一些不确定性,即柄部锥度是否可以与新的陶瓷股骨头重复使用,特别是当柄部固定良好时。陶瓷是一种脆性材料。即使男性柄部锥度有微小损伤,也可能导致应力集中,从而使新的陶瓷股骨头过早失效。因此,制造商严格禁止将柄部锥度用于重复使用的陶瓷股骨头。本研究的目的是确定组装到重复使用的男性柄部锥度上的陶瓷股骨头的断裂强度,这些柄部锥度之前经历过股骨头断裂。使用五个12/14 Ti6Al4V男性锥度和15个Al2O3陶瓷股骨头(BIOLOX forte(®);∅28 mm,L)进行连续三次断裂试验。在每次断裂试验前后,对所有部件进行目视检查并分析表面几何形状。平均断裂力(52.5 kN)并未随着锥度重复使用次数的增加而降低(p≥0.77),但范围从最初的4.1 kN显著增加到第一次重复使用时的31.8 kN和第二次重复使用时的52.6 kN,这是因为一些部件在非常低的载荷下就失效了。目视检查不足以预测降低的失效载荷。因此,在没有金属转接套筒的情况下,陶瓷股骨头不应安装在使用过的男性锥度上。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验