污染与清洁对全髋关节置换术中模块化股骨头锥度固定强度的影响。

The Influence of Contamination and Cleaning on the Strength of Modular Head Taper Fixation in Total Hip Arthroplasty.

作者信息

Krull Annika, Morlock Michael M, Bishop Nicholas E

机构信息

Institute of Biomechanics, Hamburg University of Technology, Hamburg, Germany.

Institute of Biomechanics, Hamburg University of Technology, Hamburg, Germany; Department of Medical Technology, Faculty of Life Science, HAW Hamburg University of Applied Sciences, Hamburg, Germany.

出版信息

J Arthroplasty. 2017 Oct;32(10):3200-3205. doi: 10.1016/j.arth.2017.05.009. Epub 2017 May 13.

Abstract

BACKGROUND

Intraoperative interface contamination of modular head-stem taper junctions of hip implants can lead to poor fixation strength, causing fretting and crevice corrosion or even stem taper fracture. Careful cleaning before assembly should help to reduce these problems. The purpose of this study was to determine the effect of cleaning (with and without drying) contaminated taper interfaces on the taper fixation strength.

METHODS

Metal or ceramic heads were impacted onto titanium alloy stem tapers with cleaned or contaminated (fat or saline solution) interfaces. The same procedure was performed after cleaning and drying the contaminated interfaces. Pull-off force was used to determine the influence of contamination and cleaning on the taper strength.

RESULTS

Pull-off forces after contamination with fat were significantly lower than those for uncontaminated interfaces for both head materials. Pull-off forces after application of saline solution were not significantly different from those for uncontaminated tapers. However, a large variation in taper strength was observed, pull-off forces for cleaned and dried tapers were similar to those for uncontaminated tapers for both head materials.

CONCLUSION

Intraoperative contamination of taper interfaces may be difficult to detect but has a major influence on taper fixation strength. Cleaning of the stem taper with saline solution and drying with gauze directly before assembly allows the taper strength of the pristine components to be achieved. Not drying the taper results in a large variation in pull-off forces, emphasizing that drying is essential for sufficient and reproducible fixation strength.

摘要

背景

髋关节植入物模块化头-柄锥形连接处的术中界面污染可导致固定强度不佳,引起微动和缝隙腐蚀,甚至柄部锥形骨折。组装前仔细清洁应有助于减少这些问题。本研究的目的是确定清洁(有无干燥)受污染的锥形界面对锥形固定强度的影响。

方法

将金属或陶瓷头撞击到具有清洁或受污染(脂肪或盐溶液)界面的钛合金柄部锥形上。对受污染的界面进行清洁和干燥后,进行相同的操作。使用拔出力来确定污染和清洁对锥形强度的影响。

结果

两种头部材料在被脂肪污染后的拔出力均显著低于未受污染的界面。施加盐溶液后的拔出力与未受污染的锥形无显著差异。然而,观察到锥形强度存在很大差异,两种头部材料清洁和干燥后的锥形拔出力与未受污染的锥形相似。

结论

术中锥形界面污染可能难以检测,但对锥形固定强度有重大影响。在组装前直接用盐溶液清洁柄部锥形并用纱布干燥可实现原始部件的锥形强度。不干燥锥形会导致拔出力有很大差异,强调干燥对于获得足够且可重复的固定强度至关重要。