Wang Shengqun, Wang Jiayin, Gu Changyue, Zuo Jianlin
Department of Orthopedics, China-Japan Union Hospital of Jilin University, Changchun 130033, China.
Email:
Zhonghua Wai Ke Za Zhi. 2015 Feb;53(2):90-4.
To investigate normal bony anatomy of the glenoid rim, to measure inner glenoid rim angle and outer glenoid rim angle, and the angles for successful anchor insertion for arthroscopic labral repairs.
Twelve unpaired isolated human glenoids (6 right, 6 left) without any evidence of trauma were for studying. The glenoid specimens were scanned using 320-slice CT (Aquilion ONE), then reconstruction glenoid to a three dimensional model using materialise's interactive medical image control system (Mimics) and to obtain cross-sectional images in 6 different planes, mark the right glenoid rim with clockwise tag, the left with counterclockwise tag. Inner glenoid rim angle marked as angle α and outer glenoid rim angle marked as angle β were measured from the cross-sectional images of the glenoids at 8 positions: 2-, 3-, 4-, 5-, 6-, 7-, 8- and 9-o'clock positions. Glenoid morphology was noted for each position. Using 12 mm as radius, measured the minimum insertion angle of anchor, marked as angle γ. Normal distribution of the data was confirmed with Kolmogorov-Smirnov test. Paired t-test was performed to detect differences in the angles between two locations. Two independent samples t-test was performed to detect differences in the angles between same location of left and right. Analysis of variance (ANOVA) was performed to detect differences in the angles between right and left, and different locations of the glenoid rim.
The smallest α was at the 4-o'clock position (right 50° ± 6°, left 52° ± 9°), significant difference were seen when compared with the 6-o'clock position (t = 10.466, P = 0.000) or the 5-o'clock position (t = 3.754, P = 0.003), no significant difference exist between 4-o'clock position and 3-o'clock posion (t = 0.926, P = 0.374). The smallest β was at the 3-o'clock position (right 50° ± 6°, left 53° ± 10°), significant difference were seen when compared with the 6-o'clock position (t = 9.862, P = 0.000) or the 5-o'clock position (t = 3.634, P = 0.003), no significant difference exist between 4-o'clock position and 3-o'clock posion (t = 0.697, P = 0.501). Asymmetric morphology of the glenoid was noted with an almost straight line extending medially from the rim at the 3-o'clock position, whereas a concave morthology was noted at the 9-o'clock position. Similary at the 4- and 5-o'clock position, the scapular bony surface did not curve toward the base as markedly as it did at the corresponding posterior 8- and 7-o'clock position. Angle γ from the 3-o'clock to the 9-o'clock were 25° ± 4°, 54° ± 6°, 83° ± 4°, 119° ± 2° at right side, 23° ± 4°, 57° ± 4°, 89° ± 7°, 119° ± 4° at left side. No significant difference of any angle at the same position was noted between left and right (α:t = 0.283-1.785, P > 0.05;β:t = 0.369-2.067, P > 0.05;γ:t = 0.145-0.492, P > 0.05).
The available bone mass for the anchor insertion is found to vary depending on the position of the glenoid rim. The smallest inner and outer glenoid rim angle are at the 4- and 3-o'clock position. The minimum insertion angles of anchor differ at different position. Both rim angle and glenoid morthology for each position must be considered when selecting the ideal anchor insertion angle for Bankart repair. Meanwhile, minimum insertion angle of anchor should also be considered before anchor insertion.
研究肩胛盂边缘的正常骨质解剖结构,测量肩胛盂内缘角和外缘角,以及关节镜下盂唇修复时锚钉成功置入的角度。
选取12个无创伤迹象的未配对离体人肩胛盂(6个右侧,6个左侧)用于研究。使用320层CT(Aquilion ONE)对肩胛盂标本进行扫描,然后使用Materialise的交互式医学图像控制系统(Mimics)将肩胛盂重建为三维模型,并获取6个不同平面的横断面图像,右侧肩胛盂边缘按顺时针标记,左侧按逆时针标记。从肩胛盂在8个位置(2点、3点、4点、5点、6点、7点、8点和9点位置)的横断面图像中测量标记为α角的肩胛盂内缘角和标记为β角的肩胛盂外缘角。记录每个位置的肩胛盂形态。以12 mm为半径,测量锚钉的最小置入角度,标记为γ角。用Kolmogorov-Smirnov检验确认数据的正态分布。采用配对t检验检测两个位置之间角度的差异。采用两独立样本t检验检测左右相同位置之间角度的差异。采用方差分析(ANOVA)检测左右之间以及肩胛盂边缘不同位置之间角度的差异。
最小的α角在4点位置(右侧50°±6°,左侧52°±9°),与6点位置(t = 10.466,P = 0.000)或5点位置(t = 3.754,P = 0.003)相比有显著差异,4点位置与3点位置之间无显著差异(t = 0.926,P = 0.374)。最小的β角在3点位置(右侧50°±6°,左侧53°±10°),与6点位置(t = 9.862,P = 0.000)或5点位置(t = 3.634,P = 0.003)相比有显著差异,4点位置与3点位置之间无显著差异(t = 0.697,P = 0.501)。观察到肩胛盂形态不对称,在3点位置从边缘向内侧延伸几乎呈一条直线,而在9点位置呈凹形。同样在4点和5点位置,肩胛骨关节面不像在相应的后方8点和7点位置那样明显地向底部弯曲。右侧从3点到9点位置的γ角分别为25°±4°、54°±6°、83°±4°、119°±2°,左侧分别为23°±4°、57°±4°、89°±7°、119°±4°。左右相同位置的任何角度均无显著差异(α:t = 0.283 - 1.785,P > 0.05;β:t = 0.369 - 2.067,P > 0.05;γ:t = 0.145 - 0.492,P > 0.05)。
发现锚钉置入的可用骨量因肩胛盂边缘的位置而异。最小的肩胛盂内缘角和外缘角分别在4点和3点位置。锚钉的最小置入角度在不同位置有所不同。在为Bankart修复选择理想的锚钉置入角度时,必须考虑每个位置的边缘角度和肩胛盂形态。同时,在置入锚钉前也应考虑锚钉的最小置入角度。