Eur Radiol. 2015 Dec;25(12):3638-47. doi: 10.1007/s00330-015-3759-3. Epub 2015 Apr 28.
To investigate the inter-rater reliability and agreement of the automated breast volume scanner (ABVS) and the diagnostic accuracy for differentiating malignant and benign lesions. The overall aim was to find out if the ABVS is applicable to daily clinical practice.
Qualifying studies were retrieved from Pubmed, EMBASE, Cochrane Library, Biosis Preview, CBM disc and by manual search and reference lists up to 30 September 2014. Pooled sensitivity and specificity of ABVS were calculated and summary receiver operating characteristic curves were drawn.
Thirteen studies were included in the meta-analysis of diagnostic accuracy and seven studies were included in the systematic review of inter-rater reliability/agreement of ABVS. For 'diagnostic accuracy', the pooled values of sensitivity, specificity, positive and negative likelihood ratios, and diagnostic odds ratio were 92 % (95 % CI 89.9-93.8), 84.9 % (82.4-87.0), 6.172 (4.364-8.730), 0.101 (0.075-0.136), and 72.226 (39.637-131.61), respectively. For the studies of inter-rater reliability/agreement, the quality was heterogeneous and no evidenced result can be pooled.
Sensitivity and specificity of ABVS for differentiating malignant and benign breast lesions were high. More sound studies focusing on inter-rater reliability/agreement of ABVS, which deeply affect the clinical utilization and generalization of ABVS, are urgently needed.
• ABVS has high sensitivity and specificity in differentiating malignant and benign breast lesions. • The quality of published inter-rater reliability studies is heterogeneous. • Empirical evidence concerning the inter-rater reliability/agreement for the ABVS is rare. • Comparison studies on ABVS and other medical imaging examinations are warranted.
研究自动乳腺容积扫描仪(ABVS)在鉴别良恶性病变方面的观察者间可靠性和准确性,探讨其是否适用于日常临床实践。
检索 Pubmed、EMBASE、Cochrane 图书馆、Biosis Preview、CBM disc 及手工检索和参考文献,检索时间截止至 2014 年 9 月 30 日。计算 ABVS 的合并敏感度和特异度,并绘制汇总受试者工作特征曲线。
共纳入 13 项关于诊断准确性的荟萃分析研究和 7 项关于 ABVS 观察者间可靠性/一致性的系统评价研究。“诊断准确性”方面,ABVS 的合并敏感度、特异度、阳性似然比、阴性似然比和诊断比值比分别为 92 %(95 % CI 89.9-93.8)、84.9 %(82.4-87.0)、6.172(4.364-8.730)、0.101(0.075-0.136)和 72.226(39.637-131.61)。关于观察者间可靠性/一致性的研究,质量参差不齐,无法进行合并分析。
ABVS 对鉴别乳腺良恶性病变具有较高的敏感度和特异度。需要更多关于 ABVS 观察者间可靠性/一致性的研究,这将直接影响 ABVS 的临床应用和推广。
• ABVS 对鉴别乳腺良恶性病变具有较高的敏感度和特异度。• 已发表的观察者间可靠性研究质量参差不齐。• 关于 ABVS 观察者间可靠性/一致性的证据罕见。• 需要比较 ABVS 与其他医学影像学检查的研究。