Salvadó M, Cros M, Joemai R M S, Calzado A, Geleijns J
Faculty of Medicine and Health Sciences, Universitat Rovira i Virgili, C/Sant Llorenç, 21, 43201 Reus, Spain.
Faculty of Medicine and Health Sciences, Universitat Rovira i Virgili, C/Sant Llorenç, 21, 43201 Reus, Spain.
Phys Med. 2015 Jul;31(5):452-62. doi: 10.1016/j.ejmp.2015.04.007. Epub 2015 May 8.
The purpose of this study was to develop and validate a Monte Carlo (MC) simulation tool for patient dose assessment for a 320 detector-row CT scanner, based on the recommendations of International Commission on Radiological Protection (ICRP). Additionally, the simulation was applied on four clinical acquisition protocols, with and without automatic tube current modulation (TCM).
The MC simulation was based on EGS4 code and was developed specifically for a 320 detector-row cone-beam CT scanner. The ICRP adult reference phantoms were used as patient models. Dose measurements were performed free-in-air and also in four CTDI phantoms: 150 mm and 350 mm long CT head and CT body phantoms. The MC program was validated by comparing simulations results with these actual measurements acquired under the same conditions. The measurements agreed with the simulations across all conditions within 5%. Patient dose assessment was performed for four clinical axial acquisitions using the ICRP adult reference phantoms, one of them using TCM.
The results were nearly always lower than those obtained from other dose calculator tools or published in other studies, which were obtained using mathematical phantoms in different CT systems. For the protocol with TCM organ doses were reduced by between 28 and 36%, compared to the results obtained using a fixed mA value.
The developed simulation program provides a useful tool for assessing doses in a 320 detector-row cone-beam CT scanner using ICRP adult reference computational phantoms and is ready to be applied to more complex protocols.
本研究的目的是根据国际放射防护委员会(ICRP)的建议,开发并验证一种用于320排探测器CT扫描仪患者剂量评估的蒙特卡罗(MC)模拟工具。此外,该模拟应用于四种临床采集协议,包括有无自动管电流调制(TCM)的情况。
MC模拟基于EGS4代码,专为320排探测器锥形束CT扫描仪开发。使用ICRP成人参考体模作为患者模型。在空气中以及在四个CTDI体模中进行剂量测量:150毫米和350毫米长的CT头部和CT身体体模。通过将模拟结果与在相同条件下获得的实际测量结果进行比较,对MC程序进行了验证。在所有条件下,测量结果与模拟结果的偏差在5%以内。使用ICRP成人参考体模对四种临床轴向采集进行患者剂量评估,其中一种使用TCM。
结果几乎总是低于从其他剂量计算工具获得的结果或其他研究中发表的结果,其他研究是在不同CT系统中使用数学体模获得的。与使用固定毫安值获得的结果相比,对于采用TCM的协议,器官剂量降低了28%至36%。
所开发的模拟程序为使用ICRP成人参考计算体模评估320排探测器锥形束CT扫描仪中的剂量提供了一个有用的工具,并准备应用于更复杂的协议。