Diagnostic Accuracy of Charge-coupled Device Sensor and Photostimulable Phosphor Plate Receptor in the Detection of External Root Resorption In Vitro.

作者信息

Sakhdari Shirin, Khalilak Zohreh, Najafi Elham, Cheraghi Rezvaneh

机构信息

Assistant Professor, Department of Oral & Maxillofacial Radiology, Maxillofacial Research Center,Islamic Azad University, Dental Branch, Tehran- Iran.

Associate Professor, Department of Endodontics, Islamic Azad University, Dental Branch, Tehran- Iran.

出版信息

J Dent Res Dent Clin Dent Prospects. 2015 Winter;9(1):18-22. doi: 10.15171/joddd.2015.004. Epub 2015 Mar 4.

Abstract

Background and aims. Early diagnosis of external root resorption is important for accurate treatment. The purpose of this study was to compare the efficacy of a charge-coupled device (CCD) sensor and a photostimulable phosphor (PSP) plate receptor in the diagnosis of artificial external root resorption. Materials and methods. In this diagnostic in-vitro study, 40 maxillary incisors were mounted in a segment of dry bone and preliminary radiographs were obtained using CCD and PSP sensors. Artificial resorption cavities were produced on the middle-third in half of the samples and on the cervical-third in the other half on the buccal root surfaces. Radiographs were repeated and images were evaluated. Data were statistically analyzed using chi-square and diagnostic tests. Results. There were no significant differences between the two sensors in the sensitivity (p=0.08 and 0.06) and specificity (p=0.13) for the diagnosis of resorption in both root areas. The overall accuracy of CCD was higher than PSP sensor; however, the difference was not statistically significance (p>0.05). Conclusion. CCD and PSP sensors chosen for the present study produced similar results in diagnosing simulated external root resorption.

摘要

背景与目的。早期诊断牙根外吸收对于准确治疗很重要。本研究的目的是比较电荷耦合器件(CCD)传感器和光激励荧光板(PSP)感受器在诊断人工牙根外吸收中的效果。材料与方法。在这项体外诊断研究中,将40颗上颌切牙安装在一段干燥骨中,并使用CCD和PSP传感器获取初步X线片。在一半样本的颊侧牙根表面中三分之一处制作人工吸收腔,在另一半样本的颊侧牙根表面颈三分之一处制作人工吸收腔。重复拍摄X线片并对图像进行评估。使用卡方检验和诊断试验对数据进行统计学分析。结果。在两个牙根区域的吸收诊断中,两种传感器在敏感性(p=0.08和0.06)和特异性(p=0.13)方面均无显著差异。CCD的总体准确性高于PSP传感器;然而,差异无统计学意义(p>0.05)。结论。本研究中选择的CCD和PSP传感器在诊断模拟牙根外吸收时产生了相似的结果。