眼表面干涉仪测量脂质层厚度的观察者内和观察者间可重复性。

Intra-observer and inter-observer repeatability of ocular surface interferometer in measuring lipid layer thickness.

机构信息

Yong Loo Lin School of Medicine, National University of Singapore, Singapore, Singapore.

Singapore National Eye Center, 11 Third Hospital Avenue, 168751, Singapore, Singapore.

出版信息

BMC Ophthalmol. 2015 May 15;15:53. doi: 10.1186/s12886-015-0036-9.

Abstract

BACKGROUND

Tear lipid morphology is important for normal tear function. Recently, there have been clinical studies using interferometry to assess lipid layer thickness (LLT). The aim of the study is to examine the repeatability of a commercially available interferometer.

METHODS

Two observers measured LLT in twenty Asian subjects (20 eyes) using an interferometer (LipiView® ocular surface interferometer, TearScience Inc, Morrisville, NC). Dry eye symptoms, tear break up time (TBUT) and corneal fluorescein staining were also prospectively evaluated.

RESULTS

Data for 20 participants are presented for either right or left eye (randomly selected). The mean LLT ± standard deviation of these participants was 53.53 ± 14.59 nm. When a single observer repeated the imaging on the same day, the coefficient of repeatability was 16 nm and the 95% limits of agreement were between -11 nm and 18 nm. When a different observer repeated the scan, the coefficient of repeatability was 13 nm and limits of agreement were -9 nm and 16 nm. LLT was not significantly associated with TBUT, presence of any corneal staining in any corneal zones, or symptomatic status.

CONCLUSION

With the repeatability of measurements being known, the significance of LLT changes measured by this interferometer may be better interpreted. In this small Asian study, the LLT was lower than previously reported studies.

摘要

背景

泪液脂质形态对于正常的泪液功能很重要。最近,已有临床研究使用干涉仪来评估脂质层厚度(LLT)。本研究旨在检验一种市售干涉仪的可重复性。

方法

两名观察者使用干涉仪(LipiView®眼表面干涉仪,TearScience Inc,Morrisville,NC)对二十名亚洲受试者(二十只眼)的 LLT 进行了测量。还前瞻性评估了干眼症状、泪膜破裂时间(TBUT)和角膜荧光素染色。

结果

为右眼或左眼(随机选择)呈现了 20 名参与者的数据。这些参与者的平均 LLT±标准偏差为 53.53±14.59nm。当一名观察者在同一天重复成像时,可重复性系数为 16nm,95%一致性界限在-11nm 和 18nm 之间。当不同的观察者重复扫描时,可重复性系数为 13nm,一致性界限在-9nm 和 16nm 之间。LLT 与 TBUT 、任何角膜区域的任何角膜染色的存在或症状状态均无显著相关性。

结论

在了解了测量的可重复性后,可能会更好地解释该干涉仪测量的 LLT 变化的意义。在这项小型亚洲研究中,LLT 低于先前报道的研究。