Minami Shujiro B, Takegoshi Hideki, Shinjo Yukiko, Enomoto Chieko, Kaga Kimitaka
National Institute of Sensory Organs, National Tokyo Medical Center , Tokyo , Japan.
Acta Otolaryngol. 2015;135(10):1007-15. doi: 10.3109/00016489.2015.1048377. Epub 2015 Jun 10.
EABR is a reliable and effective way of objectively confirming device function and implant-responsiveness of the peripheral auditory neurons up to the level of the brainstem in cases of inner ear malformation.
To investigate the usefulness of measuring the intra-operative electrically evoked compound action potential (ECAP) and electrically evoked auditory brainstem response (EABR) in patients with and without congenital inner ear anomalies during cochlear implantation.
Thirty-eight consecutive children (40 ears) aged 5 or younger with congenital profound hearing loss. Twenty-four (25 ears) lacked congenital inner ear anomalies. The 14 patients (15 ears) with a malformation had common cavities (four ears), incomplete partition type I (three ears), cochlea hypoplasia type III (three ears), enlarged vestibular aqueduct (four ears), and cochlear nerve canal stenosis (one ear). Main outcome measures are ECAP and EABR responses.
Of the 25 ears lacking any malformation, 21, three, and one showed 'Good', 'Variable', and 'No' ECAP responses, respectively, and 24 and one showed 'Good' and 'Variable' intra-cochlear responses, respectively. Of the 15 ears with a malformation, two showed 'Good' ECAP responses, nine had 'Variable' ECAP responses, and four showed 'No' ECAP responses. Moreover, five showed 'Good' EABR responses and 10 showed 'Variable' EABR responses.
在先天性内耳畸形病例中,电刺激听性脑干反应(EABR)是客观确认装置功能以及外周听觉神经元直至脑干水平植入反应性的一种可靠且有效的方法。
探讨在人工耳蜗植入术中测量有或无先天性内耳异常患者的术中电诱发复合动作电位(ECAP)和电刺激听性脑干反应(EABR)的效用。
连续纳入38名年龄5岁及以下的先天性重度听力损失儿童(40耳)。其中24名(25耳)无先天性内耳异常。14名(15耳)有畸形的患者中,有共同腔(4耳)、不完全分隔I型(3耳)、III型耳蜗发育不全(3耳)、前庭导水管扩大(4耳)以及耳蜗神经管狭窄(1耳)。主要观察指标为ECAP和EABR反应。
在25耳无任何畸形的患者中,分别有21耳、3耳和1耳显示“良好”、“可变”和“无”ECAP反应,分别有24耳和1耳显示“良好”和“可变”的耳蜗内反应。在15耳有畸形的患者中,2耳显示“良好”ECAP反应,9耳有“可变”ECAP反应,4耳显示“无”ECAP反应。此外,5耳显示“良好”EABR反应,10耳显示“可变”EABR反应。