Hobson Maritza A, Davis Stephen D
McGill University Health Centre, Department of Medical Physics.
J Appl Clin Med Phys. 2015 Mar 8;16(2):4973. doi: 10.1120/jacmp.v16i2.4973.
While commissioning Varian's Portal Dose Image Prediction (PDIP) algorithm for portal dosimetry, an asymmetric radial response in the portal imager due to backscatter from the support arm was observed. This asymmetric response led to differences on the order of 2%-3% for simple square fields (< 20 × 20 cm2) when comparing the measured to predicted portal fluences. A separate problem was that discrepancies of up to 10% were seen in measured to predicted portal fluences at increasing off-axis distance (> 10 cm). We have modified suggested methods from the literature to provide a 1D correction for the off-axis response problem which adjusts the diagonal profile used in the portal imager calibration. This inherently cannot fix the 2D problem since the PDIP algorithm assumes a radially symmetric response and will lead to some uncertainty in portal dosimetry results. Varian has recently released generic "2D correction" files with their Portal Dosimetry Pre-configuration (PDPC) package, but no independent testing has been published. We present the comparison between QA results using the Varian correction method to results using our 1D profile correction method using the gamma passing rates with a 3%, 3 mm criterion. The average, minimum, and maximum gamma pass rates for nine fixed-field IMRT fields at gantry 0° using our profile correction method were 98.1%, 93.7%, and 99.8%, respectively, while the results using the PDPC correction method were 98.4%, 93.1%, and 99.8%. For four RapidArc fields, the average, minimum, and maximum gamma pass rates using our correction method were 99.6%, 99.4%, and 99.9%, respectively, while the results using the PDPC correction method were 99.8%, 99.5%, and 99.9%. The average gamma pass rates for both correction methods are quite similar, but both show improvement over the uncorrected results.
在调试瓦里安公司的射野剂量图像预测(PDIP)算法用于射野剂量测定时,观察到由于支撑臂的反向散射,射野成像仪中出现了不对称的径向响应。这种不对称响应导致在比较测量的和预测的射野注量时,对于简单方形射野(<20×20 cm²),差异在2%-3%左右。另一个问题是,在离轴距离增加(>10 cm)时,测量的和预测的射野注量之间出现了高达10%的差异。我们对文献中建议的方法进行了修改,以针对离轴响应问题提供一维校正,该校正会调整射野成像仪校准中使用的对角线轮廓。由于PDIP算法假定为径向对称响应,所以这本质上无法解决二维问题,并且会在射野剂量测定结果中导致一些不确定性。瓦里安公司最近随其射野剂量测定预配置(PDPC)软件包发布了通用的“二维校正”文件,但尚未发表独立测试结果。我们给出了使用瓦里安校正方法的质量保证(QA)结果与使用我们的一维轮廓校正方法的结果之间的比较,采用了3%、3 mm标准的伽马通过率。使用我们的轮廓校正方法,在机架0°时九个固定野调强放疗(IMRT)射野的平均、最小和最大伽马通过率分别为98.1%、93.7%和99.8%,而使用PDPC校正方法的结果分别为98.4%、93.1%和99.8%。对于四个容积调强弧形治疗(RapidArc)射野,使用我们的校正方法的平均、最小和最大伽马通过率分别为99.6%、99.4%和99.9%,而使用PDPC校正方法的结果分别为99.8%、99.5%和99.9%。两种校正方法的平均伽马通过率非常相似,但与未校正的结果相比都有改进。