一种用于离轴射野剂量测定误差的有效校正算法。

An effective correction algorithm for off-axis portal dosimetry errors.

作者信息

Bailey Daniel W, Kumaraswamy Lalith, Podgorsak Matthew B

机构信息

Department of Physics, State University of New York at Buffalo, Buffalo, New York 14260, USA.

出版信息

Med Phys. 2009 Sep;36(9):4089-94. doi: 10.1118/1.3187785.

Abstract

Portal dosimetric images acquired for IMRT pretreatment verification show dose errors of up to 15% near the detector edges as compared to dose predictions calculated by a treatment planning system for these off-axis regions. A method is proposed to account for these off-axis effects by precisely correcting the off-axis output factors, which calibrate the imager for absolute dose. Using this method, agreement between the predicted and the measured doses improves by up to 15% for fields near the detector edges, resulting in passing rate improvements of as much as 60% for gamma evaluation of 3 mm, 3% within the collimator jaws.