Lussani Fernando Cesar, Vescovi Rafael Ferreira da Costa, de Souza Thaís Diniz, Leite Carlos A P, Giles Carlos
Instituto de Física "Gleb Wataghin," Universidade Estadual de Campinas, Campinas, São Paulo 13083-859, Brazil.
Rev Sci Instrum. 2015 Jun;86(6):063705. doi: 10.1063/1.4922607.
An x-ray microtomography station implemented at the X-ray Applied Crystallography Laboratory of the State University of Campinas is described. The station is based on a propagation based phase contrast imaging setup with a microfocus source and digital x-ray area detectors. Due to its simplicity, this setup is ideal for fast, high resolution imaging and microtomography of small biological specimens and materials research samples. It can also be coupled to gratings to use and develop new techniques as the harmonic spatial coherent imaging, which allow scattering contrast imaging. Details of the experimental setup, equipment, and software integration are described. Test microtomography for setup commissioning and characterization is shown. We conclude that phase contrast enhanced x-ray imaging and microtomography with resolution below 5 μm voxel size are possible and data sets as wide as 2000 × 2000 × 2000 voxels are obtained with this instrumentation.
描述了在坎皮纳斯州立大学的X射线应用晶体学实验室实施的一个X射线显微断层扫描站。该站基于一个基于传播的相衬成像装置,配备微焦点源和数字X射线面积探测器。由于其简单性,这种装置非常适合对小型生物标本和材料研究样品进行快速、高分辨率成像和显微断层扫描。它还可以与光栅耦合,以使用和开发新技术,如谐波空间相干成像,从而实现散射对比成像。描述了实验装置、设备和软件集成的细节。展示了用于装置调试和表征的测试显微断层扫描。我们得出结论,使用该仪器可以实现相衬增强X射线成像和体素尺寸低于5μm的显微断层扫描,并且可以获得高达2000×2000×2000体素的数据集。