Huang Eddie H, Copp Steven N, Bugbee William D
University of Texas Health Science Center, Houston, Texas.
Division of Orthopaedic Surgery, Scripps Clinic, La Jolla, California.
J Arthroplasty. 2015 Nov;30(11):1906-10. doi: 10.1016/j.arth.2015.05.055. Epub 2015 Jun 6.
Restoration of mechanical axis in total knee arthroplasty (TKA) is correlated with improved implant survivorship. We assessed the accuracy and required surgical time using a hand-held accelerometer-based navigation system for TKA. Data collected on 53 patients included assembly, resection, and tourniquet times. Implant alignment and mechanical axis were measured on radiographs. Femoral alignment was 0.29° ± 2.2° varus. Tibial alignment was 0.09° ± 1.4° valgus. Postoperative mechanical axis was 0.2° ± 2.1° varus. Malalignment rates for the femur, tibia, and axis were 13%, 3.8%, and 17%, respectively. Average time for pinning and navigating was 3.6 minutes for the femur and 2.6 minutes for the tibia; mean tourniquet time was 62 minutes. This navigation system accurately re-established mechanical axis without increasing surgical time.
全膝关节置换术(TKA)中机械轴的恢复与植入物存活率的提高相关。我们使用基于手持加速度计的导航系统对TKA的准确性和所需手术时间进行了评估。收集的53例患者的数据包括组装、切除和止血带使用时间。通过X线片测量植入物对线和机械轴。股骨对线为内翻0.29°±2.2°。胫骨对线为外翻0.09°±1.4°。术后机械轴为内翻0.2°±2.1°。股骨、胫骨和轴的对线不良率分别为13%、3.8%和17%。股骨的打钉和导航平均时间为3.6分钟,胫骨为2.6分钟;平均止血带使用时间为62分钟。该导航系统可准确重建机械轴,且不增加手术时间。