van Norren D, van de Kraats J
Royal Netherlands Eye Hospital, Utrecht.
Vision Res. 1989;29(12):1825-30. doi: 10.1016/0042-6989(89)90163-6.
We describe a novel use of the Scanning Laser Ophthalmoscope (SLO), viz. as an imaging retinal densitometer. In our SLO a helium-neon or an argon laser beam is moved in a raster pattern over the retina; the reflected light is descanned (confocal SLO) and collected by a photomultiplier. Images of the fundus subtending 22 by 18 deg are displayed on a TV monitor. Single frames taken with 514 nm light were stored in a computer in arrays of 256 by 256 pixels and density differences between dark adapted and bleached images were calculated. With a full bleach density differences of about 0.35 were found in the center of the fovea; at retinal eccentricities of 15-20 deg we found 0.15. After selective bleaching with 633 nm light substantial density differences were only seen in the foveal area. We conclude that the confocal SLO is a very suitable instrument for imaging fundus reflectometry.