Lenk Gabriel, Sandkvist Sören, Pohanka Anton, Stemme Göran, Beck Olof, Roxhed Niclas
Department for Micro & Nanosystems, KTH Royal Institute of Technology, Osquldas väg 10, 10044 Stockholm, Sweden.
Department of Clinical Pharmacology, Karolinska University Hospital, 14152 Stockholm, Sweden.
Bioanalysis. 2015;7(16):2085-94. doi: 10.4155/bio.15.134.
DBS samples collected from a fingerprick typically vary in volume and homogeneity and hence make an accurate quantitative analysis of DBS samples difficult.
We report a prototype which first defines a precise liquid volume and subsequently stores it to a conventional DBS matrix. Liquid volumes of 2.2 µl ± 7.1% (n = 21) for deionized water and 6.1 µl ± 8.8% (n = 15) for whole blood have been successfully metered and stored in DBS paper.
The new method of collecting a defined volume of blood by DBS sampling has the potential to reduce assay bias for the quantitative evaluation of DBS samples while maintaining the simplicity of conventional DBS sampling.
通过手指采血收集的干血斑样本在体积和均一性方面通常存在差异,因此难以对干血斑样本进行准确的定量分析。
我们报告了一种原型,该原型首先定义精确的液体体积,随后将其存储到传统的干血斑基质中。已成功计量并将2.2微升±7.1%(n = 21)的去离子水和6.1微升±8.8%(n = 15)的全血液体体积存储在干血斑滤纸上。
通过干血斑采样收集确定体积血液的新方法,在保持传统干血斑采样简便性的同时,有可能减少干血斑样本定量评估中的检测偏差。