Venema Leonie H, Sharma Ajay S, Simons Antoine P, Bekers Otto, Weerwind Patrick W
J Extra Corpor Technol. 2014 Sep;46(3):212-6.
Hemolysis is a well-known phenomenon during cardiovascular surgery and generally attributed to cardiopulmonary bypass, particularly when using high-resistant oxygenators. This study aimed at investigating whether transoxygenator pressure drop can be considered an independent factor of hemolysis. Additionally, intraoxygenator blood distribution and shear stress were assessed. A low-resistant (LR, n = 3), a moderate-resistant (MR, n = 3), and a high-resistant (HR, n = 3) clinically used membrane oxygenator were tested in vitro using a roller pump and freshly drawn heparinized porcine blood. Flow rates were set to 2 and 4 L/min and maximum flow compliant to the oxygenator type for 1 hour each. As a control, the oxygenator was excluded from the system. Blood samples were taken every 30 minutes for plasma-free hemoglobin assay and transoxygenator pressure was measured inline. Intraoxygenator blood distribution was assessed using an ultrasound dilution technique. Despite the relatively broad spectrum of pressure drop and resultant transoxygenator pressure drops (LR: 14-41 mmHg, MR: 29-115 mmHg, HR: 77-284 mmHg, respectively), no significant association (R2 = .074, p = .22) was found with the normalized index of hemolysis. The shear stress of each oxygenator at maximum flow rate amounted to 3.0 N/m2 (LR), 5.7 N/m2 (MR), and 8.4 N/m2 (HR), respectively. Analysis of blood flow distribution curves (kurtosis and skewness) revealed intraoxygenator blood flow distribution to become more homogeneous when blood flow rates increased. Contemporary oxygenators were shown not to be a predominant factor for red blood cell damage.
溶血是心血管手术中一种众所周知的现象,通常归因于体外循环,尤其是在使用高阻力氧合器时。本研究旨在调查跨氧合器压降是否可被视为溶血的一个独立因素。此外,还评估了氧合器内血液分布和剪切应力。使用滚压泵和新鲜抽取的肝素化猪血,对三种临床上使用的低阻力(LR,n = 3)、中阻力(MR,n = 3)和高阻力(HR,n = 3)膜式氧合器进行了体外测试。流速设定为2 L/min和4 L/min,并在每种流速下以符合氧合器类型的最大流速运行1小时。作为对照,将氧合器从系统中排除。每30分钟采集一次血样进行无血浆血红蛋白测定,并在线测量跨氧合器压力。使用超声稀释技术评估氧合器内血液分布。尽管压降范围相对较宽,导致跨氧合器压降分别为(LR:14 - 41 mmHg,MR:29 - 115 mmHg,HR:77 - 284 mmHg),但未发现与溶血标准化指数有显著关联(R2 = 0.074,p = 0.22)。每种氧合器在最大流速下的剪切应力分别为3.0 N/m2(LR)、5.7 N/m2(MR)和8.4 N/m2(HR)。血流分布曲线(峰度和偏度)分析表明,当血流速度增加时,氧合器内血流分布变得更加均匀。现代氧合器并非红细胞损伤的主要因素。