De Somer F
Heart Centre, University Hospital Ghent, B-9000 Ghent, Belgium.
Perfusion. 2013 Jul;28(4):280-5. doi: 10.1177/0267659113483803. Epub 2013 Apr 25.
Based on their design, all membrane oxygenators generate a certain resistance to flow. In clinical practice, this resistance is calculated by measuring both blood flow and the pressure drop over the oxygenator. Historically, some designs, such as the Kolobow spiral coil oxygenator and the Cobe flat sheet oxygenator, had quite a high pressure drop, but were, nevertheless, considered very haemocompatible. Today, both medium and low pressure drop oxygenators are commercially available. Based on physics and the existing literature, this review aims to investigate whether pressure drop by itself can be considered an independent factor of haemolysis.
基于其设计,所有膜式氧合器都会对血流产生一定阻力。在临床实践中,这种阻力是通过测量血流和氧合器两端的压降来计算的。从历史上看,一些设计,如科洛博夫螺旋盘管式氧合器和科比平板式氧合器,具有相当高的压降,但仍被认为具有很好的血液相容性。如今,中低压降的氧合器在市场上均有销售。基于物理学原理和现有文献,本综述旨在研究压降本身是否可被视为溶血的一个独立因素。