Optimization of SPECT-CT Hybrid Imaging Using Iterative Image Reconstruction for Low-Dose CT: A Phantom Study.

作者信息

Grosser Oliver S, Kupitz Dennis, Ruf Juri, Czuczwara Damian, Steffen Ingo G, Furth Christian, Thormann Markus, Loewenthal David, Ricke Jens, Amthauer Holger

机构信息

Department of Radiology and Nuclear Medicine, University Hospital Magdeburg, Magdeburg, Germany.

出版信息

PLoS One. 2015 Sep 21;10(9):e0138658. doi: 10.1371/journal.pone.0138658. eCollection 2015.

Abstract

BACKGROUND

Hybrid imaging combines nuclear medicine imaging such as single photon emission computed tomography (SPECT) or positron emission tomography (PET) with computed tomography (CT). Through this hybrid design, scanned patients accumulate radiation exposure from both applications. Imaging modalities have been the subject of long-term optimization efforts, focusing on diagnostic applications. It was the aim of this study to investigate the influence of an iterative CT image reconstruction algorithm (ASIR) on the image quality of the low-dose CT images.

METHODOLOGY/PRINCIPAL FINDINGS: Examinations were performed with a SPECT-CT scanner with standardized CT and SPECT-phantom geometries and CT protocols with systematically reduced X-ray tube currents. Analyses included image quality with respect to photon flux. Results were compared to the standard FBP reconstructed images. The general impact of the CT-based attenuation maps used during SPECT reconstruction was examined for two SPECT phantoms. Using ASIR for image reconstructions, image noise was reduced compared to FBP reconstructions for the same X-ray tube current. The Hounsfield unit (HU) values reconstructed by ASIR were correlated to the FBP HU values(R2 ≥ 0.88) and the contrast-to-noise ratio (CNR) was improved by ASIR. However, for a phantom with increased attenuation, the HU values shifted for low X-ray tube currents I ≤ 60 mA (p ≤ 0.04). In addition, the shift of the HU values was observed within the attenuation corrected SPECT images for very low X-ray tube currents (I ≤ 20 mA, p ≤ 0.001).

CONCLUSION/SIGNIFICANCE: In general, the decrease in X-ray tube current up to 30 mA in combination with ASIR led to a reduction of CT-related radiation exposure without a significant decrease in image quality.

摘要

背景

混合成像将核医学成像(如单光子发射计算机断层扫描(SPECT)或正电子发射断层扫描(PET))与计算机断层扫描(CT)相结合。通过这种混合设计,接受扫描的患者会累积来自这两种检查的辐射暴露。成像模式一直是长期优化努力的主题,重点在于诊断应用。本研究的目的是调查迭代CT图像重建算法(ASIR)对低剂量CT图像质量的影响。

方法/主要发现:使用具有标准化CT和SPECT体模几何形状以及系统降低X射线管电流的CT协议的SPECT-CT扫描仪进行检查。分析包括与光子通量相关的图像质量。将结果与标准滤波反投影(FBP)重建图像进行比较。针对两个SPECT体模,检查了SPECT重建期间使用的基于CT的衰减图的总体影响。使用ASIR进行图像重建时,与相同X射线管电流下的FBP重建相比,图像噪声降低。ASIR重建的亨氏单位(HU)值与FBP的HU值相关(R2≥0.88),并且ASIR提高了对比度噪声比(CNR)。然而,对于衰减增加的体模,在低X射线管电流I≤60 mA时HU值发生了偏移(p≤0.04)。此外,在极低X射线管电流(I≤20 mA,p≤0.001)的衰减校正SPECT图像中也观察到了HU值的偏移。

结论/意义:一般来说,将X射线管电流降低至30 mA并结合ASIR可减少与CT相关的辐射暴露且图像质量无显著下降。