Suppr超能文献

一种新型平板电脑应用程序用于术中失血的实时监测

Real Time Intraoperative Monitoring of Blood Loss with a Novel Tablet Application.

作者信息

Sharareh Behnam, Woolwine Spencer, Satish Siddarth, Abraham Peter, Schwarzkopf Ran

机构信息

Department of Orthopaedic Surgery, Joint Replacement Service, University of California, Irvine, Orange, CA, USA.

Gauss Surgical, Inc. Los Altos, CA, USA.

出版信息

Open Orthop J. 2015 Aug 31;9:422-6. doi: 10.2174/1874325001509010422. eCollection 2015.

Abstract

INTRODUCTION

Real-time monitoring of blood loss is critical in fluid management. Visual estimation remains the standard of care in estimating blood loss, yet is demonstrably inaccurate. Photometric analysis, which is the referenced "gold-standard" for measuring blood loss, is both time-consuming and costly. The purpose of this study was to evaluate the efficacy of a novel tablet-monitoring device for measurement of Hb loss during orthopaedic procedures.

METHODS

This is a prospective study of 50 patients in a consecutive series of joint arthroplasty cases. The novel System with Feature Extraction Technology was used to measure the amount of Hb contained within surgical sponges intra-operatively. The system's measures were then compared with those obtained via gravimetric method and photometric analysis. Accuracy was evaluated using linear regression and Bland-Altman analysis.

RESULTS

Our results showed a significant positive correlation between Triton tablet system and photometric analysis with respect to intra-operative hemoglobin and blood loss at 0.92 and 0.91, respectively.

DISCUSSION

This novel system can accurately determine Hb loss contained within surgical sponges. We believe that this user-friendly software can be used for measurement of total intraoperative blood loss and thus aid in a more accurate fluid management protocols during orthopaedic surgical procedures.

摘要

引言

失血的实时监测在液体管理中至关重要。视觉估计仍然是估计失血量的护理标准,但明显不准确。光度分析作为测量失血量的参考“金标准”,既耗时又昂贵。本研究的目的是评估一种新型平板电脑监测设备在骨科手术中测量血红蛋白损失的有效性。

方法

这是一项对50例连续关节置换病例患者的前瞻性研究。采用具有特征提取技术的新型系统在术中测量手术海绵内所含血红蛋白的量。然后将该系统的测量结果与通过重量法和光度分析获得的结果进行比较。使用线性回归和布兰德-奥特曼分析评估准确性。

结果

我们的结果显示,Triton平板电脑系统与光度分析在术中血红蛋白和失血量方面分别存在显著正相关,相关系数分别为0.92和0.91。

讨论

这种新型系统可以准确测定手术海绵内的血红蛋白损失。我们认为,这种用户友好的软件可用于测量术中总失血量,从而有助于在骨科手术过程中制定更准确的液体管理方案。