Pibida L, Mille M, Norman B
National Institute of Standards and Technology, Gaithersburg, MD 20899.
National Institute of Standards and Technology, Gaithersburg, MD 20899 ; Nuclear Engineering and Engineering Physics Program, Rensselaer Polytechnic Institute, Troy, NY 12180.
J Res Natl Inst Stand Technol. 2013 May 28;118:292-300. doi: 10.6028/jres.118.014. eCollection 2013.
Several measurements and calculations were performed to illustrate the differences that can be observed in the determination of exposure rate or ambient dose equivalent rate used for testing radiation detection systems against consensus standards. The large variations observed support our recommendation that better consistency in the test radiation fields can be achieved by specifying the source activity and testing distance instead of the field strength.