Berg A, Gerok W, Keul J
Clin Chim Acta. 1978 May 16;86(1):95-100. doi: 10.1016/0009-8981(78)90463-1.
L-Alanine was measured by an enzymic micromethod in serum after treatment with perchloric acid, as well as after ultrafiltration through collodium membrane filters. Serum L-alanine was also determined by an automated column chromatographic technique. Using the enzymic method, spuriously increased L-alanine levels could be obtained due to the absorbent reaction between hydrazine and NAD-containing reagents. It could be shown that this absorbent reaction depends on the pH value of the test solution. In contrast to supernatant diluted by perchloric acid, ultrafiltrates gave L-alanine data equivalent to those by column chromatography, since conditions of reaction time and pH value could be kept identical for blank, standard and ultrafiltrate solutions.
用高氯酸处理后以及通过火棉胶膜过滤器超滤后的血清中的L-丙氨酸,采用酶微量法进行测定。血清L-丙氨酸也通过自动柱色谱技术进行测定。使用酶法时,由于肼与含NAD试剂之间的吸收反应,可能会得到假性升高的L-丙氨酸水平。可以证明,这种吸收反应取决于测试溶液的pH值。与用高氯酸稀释的上清液相比,超滤液给出的L-丙氨酸数据与柱色谱法得到的数据相当,因为空白、标准和超滤液溶液的反应时间和pH值条件可以保持相同。